COLLEAGUE 3 VOLUMETRIC INFUSION PUMP
Report
- Report Number
- 6000001-2007-04807
- Event Type
- Malfunction
- Date Received
- September 17, 2008
- Date of Event
- February 1, 2007
- Report Date
- February 6, 2007
- Manufacturer
- BAXTER HEALTHCARE PTE. LTD.
- Product Code
- FRN
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AZ, US
- Reporter Occupation
- OTHER
Narratives
THIS REPORT IS BEING RESUBMITTED IN ACCORDANCE WITH INSTRUCTIONS FROM FDA TO ADDRESS A MANUFACTURER REPORT SEQUENCE NUMBER ISSUE THAT OCCURRED WHEN THIS MDR OR SUPPLEMENT WAS ORIGINALLY SUBMITTED TO THE FDA ON MAR 05 2007. EVALUATION SUMMARY:THE REPORTED CONDITION OF THE FLOW RATE TOO HIGH ON CHANNELS A, B AND C COULD NOT BE CONFIRMED DURING SERVICE EVALUATION. A PUMP HEAD MECHANISM ACCURACY TEST WAS CONDUCTED ON CHANNELS A AND C AND BOTH PASSED. A PUMP HEAD MECHANISM ACCURACY TEST WAS ALSO CONDUCTED ON CHANNEL B AND UNDER INFUSION OCCURRED. THE PUMP HEAD MECHANISM ON CHANNEL B WAS REPLACED. NO CORRECTIONS TO CHANNELS A AND C WERE MADE AS THE ACCURACY TESTS PASSED.
THE FACILITY REPORTED A PUMP IN WHICH THE CHANNEL A, B AND C FLOW RATES WERE TOO HIGH. THIS PROBLEM WAS IDENTIFIED DURING BIO-MED TESTING. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION NECESSARY ACCORDING TO THE HOSPITAL REPRESENTATIVE. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | COLLEAGUE 3 VOLUMETRIC INFUSION PUMP | INFUSION PUMP | FRN | BAXTER HEALTHCARE PTE. LTD. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |