STEAM PAC STANDARD BOXED
Report
- Report Number
- 1022819-2008-00256
- Event Type
- Injury
- Date Received
- September 15, 2008
- Date of Event
- August 20, 2007
- Report Date
- August 20, 2008
- Manufacturer
- CHATTANOOGA GROUP
- Product Code
- IMA
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TN, US
- Reporter Occupation
- OTHER
Narratives
THE CLINIC COULD NOT DETERMINE THE MODEL NUMBER OF THE DEVICE. THE CLINIC DISPOSED OF THE DEVICE. THE DEVICE WAS NOT RETURNED TO THE MANUFACTURE FOR EVALUATION. THE MANUFACTURE COULD NOT DETERMINE THE ROOT CAUSE AND/OR CONDITIONS THAT LEAD TO THE DEVICE'S REPORTED FAILURE.
A FEMALE PATIENT RECEIVED A BURN DURING THE USE OF A HOT PACK. THE PATIENT WAS BEING TREATED ON THE RIGHT HIP FOR HIP PAIN. THE PATIENT WAS PLACED ON THEIR LEFT SIDE FOR THE TREATMENT. THE CLINICIAN APPLIED THE HOT PAC TO THE PATIENT'S RIGHT HIP. THE PACK WAS WRAPPED IN SIX LAYERS OF TOWELING AND APPLIED TO THE PATIENT. THE PATIENT INDICATED TO THE CLINICIAN THAT THEY FELT A HOT AREA ON THEIR LEFT SIDE. THE CLINICIAN THAT THEY FELT A HOT AREA ON THEIR LEFT SIDE. THE CLINICIAN TERMINATED THE TREATMENT, AND INSPECTED THE TREATMENT AREA AND THE PATIENT'S LEFT SIDE. THE PATIENT'S LEFT SIDE HAD A RED MARK ABOUT 1 INCH IN DIAMETER. THE MARK ON THE LEFT SIDE DID APPEAR TO BE A 2ND DEGREE BURN. A PATIENT'S RIGHT SIDE HAD NO MARKINGS FROM THE APPLICATION OF THE TREATMENT. THE CLINICIAN DID RENDER FIRST AID TO THE PATIENT WITH THE USE OF ANTISEPTIC AND GAUZE. THE PATIENT DID FOLLOW UP WITH THEIR PRIMARY CARE PHYSICIAN. UPON INSPECTION OF THE HOT PACK, THE CLINICIAN NOTED THAT THE PACK SEAM HAD SPLIT. THE CLINICIAN INDICATED THAT THE SPLIT IN THE SEAM ALLOWED THE CONTENTS OF THE PACK TO LEAK ONTO THE TREATMENT SURFACE, COMING INTO CONTACT WITH THE PATIENT'S LEFT SIDE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | STEAM PAC STANDARD BOXED | IMA | CHATTANOOGA GROUP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 80 YR | Required Intervention |