FDA Adverse Event Malfunction Summary report: N

3M BAIR HUGGER WARMING UNIT 775

MDR report key: 11611922 · Received April 4, 2021

Report

Report Number
2110898-2021-00018
Event Type
Malfunction
Date Received
April 4, 2021
Date of Event
March 3, 2021
Report Date
April 4, 2021
Manufacturer
3M HEALTH CARE
Product Code
DWJ
PMA / PMN Number
K060865
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE HAS BEEN VISUALLY INSPECTED BY TECHNICAL GOVERNANCE MANAGER AT THE HOSPITAL. DEVICE WAS MOUNTED ON A MOBILE POLE-STAND. THERE WAS FAINT EVIDENCE OF A DRIP MARK DOWN THE BACK OF THE DEVICE ABOVE THE MAINS SOCKET AS WELL ON THE BASE OF THE POLE-STAND. USERS REPORTED THAT NO LIQUID INFUSION DEVICES WERE IN USE OR ATTACHED TO THE STAND AT THE TIME OF THE INCIDENT BUT MAY HAVE BEEN MOUNTED IN THE PAST. 3M¿ WILL CONTINUE TO MONITOR.

Description of Event or Problem · 1

A HOSPITAL ALLEGED AN ALARM FROM THE WALL CONTROL PANEL ALARMED AND THE CONTROL PANEL DISPLAY SHOWED AN ERROR MESSAGE OF "EARTH FAULT IN BAY 4". ALARM WAS REPORTED TO THE HOSPITAL. APPROXIMATELY TWO HOURS LATER THERE WAS A METALLIC BURNING SMELL. AFTER FIFTEEN MINUTES OF THE METALLIC SMELL A SPARK AND ORANGE FLAME CAME FROM 3M¿ BAIR HUGGER WARMING UNIT 77507 THAT WAS PLUGGED IN A PATIENT'S ROOM. THERE WAS ONE ORANGE FLAME AND THEN THE FLAME SELF-TERMINATED. THE 3M¿ BAIR HUGGER WARMING UNIT WAS TURNED OFF AT THE WALL AND DEVICE WAS UNPLUGGED. THERE WERE NO FURTHER FLAMES AND DEVICE WAS REMOVED FROM USE. NO PATIENT OR STAFF INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
509192 3M BAIR HUGGER WARMING UNIT 775 HYPER/HYPOTHERMA SYSTEM, PRODUCT CODE: DWJ DWJ 3M HEALTH CARE 775 125271

Patients

Seq Age Sex Outcome Treatment
1