3M BAIR HUGGER WARMING UNIT 775
Report
- Report Number
- 2110898-2021-00018
- Event Type
- Malfunction
- Date Received
- April 4, 2021
- Date of Event
- March 3, 2021
- Report Date
- April 4, 2021
- Manufacturer
- 3M HEALTH CARE
- Product Code
- DWJ
- PMA / PMN Number
- K060865
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
THE DEVICE HAS BEEN VISUALLY INSPECTED BY TECHNICAL GOVERNANCE MANAGER AT THE HOSPITAL. DEVICE WAS MOUNTED ON A MOBILE POLE-STAND. THERE WAS FAINT EVIDENCE OF A DRIP MARK DOWN THE BACK OF THE DEVICE ABOVE THE MAINS SOCKET AS WELL ON THE BASE OF THE POLE-STAND. USERS REPORTED THAT NO LIQUID INFUSION DEVICES WERE IN USE OR ATTACHED TO THE STAND AT THE TIME OF THE INCIDENT BUT MAY HAVE BEEN MOUNTED IN THE PAST. 3M¿ WILL CONTINUE TO MONITOR.
A HOSPITAL ALLEGED AN ALARM FROM THE WALL CONTROL PANEL ALARMED AND THE CONTROL PANEL DISPLAY SHOWED AN ERROR MESSAGE OF "EARTH FAULT IN BAY 4". ALARM WAS REPORTED TO THE HOSPITAL. APPROXIMATELY TWO HOURS LATER THERE WAS A METALLIC BURNING SMELL. AFTER FIFTEEN MINUTES OF THE METALLIC SMELL A SPARK AND ORANGE FLAME CAME FROM 3M¿ BAIR HUGGER WARMING UNIT 77507 THAT WAS PLUGGED IN A PATIENT'S ROOM. THERE WAS ONE ORANGE FLAME AND THEN THE FLAME SELF-TERMINATED. THE 3M¿ BAIR HUGGER WARMING UNIT WAS TURNED OFF AT THE WALL AND DEVICE WAS UNPLUGGED. THERE WERE NO FURTHER FLAMES AND DEVICE WAS REMOVED FROM USE. NO PATIENT OR STAFF INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 509192 | 3M BAIR HUGGER WARMING UNIT 775 | HYPER/HYPOTHERMA SYSTEM, PRODUCT CODE: DWJ | DWJ | 3M HEALTH CARE | 775 | 125271 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |