STYLE 68 SALINE FILLED BREAST IMPLANT
Report
- Report Number
- 9617229-2021-04614
- Event Type
- Injury
- Date Received
- April 2, 2021
- Date of Event
- January 28, 2021
- Report Date
- July 9, 2021
- Manufacturer
- ALLERGAN (COSTA RICA)
- Product Code
- FWM
- PMA / PMN Number
- P990074
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- PHYSICIAN
Narratives
DEVICE EVALUATION: ANALYSIS OF THE RETURNED DEVICE IDENTIFIED: CREASE FLAT, OPENING ON ANTERIOR AND YELLOW PARTICLES INNER THE SHELL. LEAK TEST AND MICROSCOPIC ANALYSIS WAS PERFORMED WHICH IDENTIFIED: SHARP OPENING ON VALVE BOND EDGE AND OBSERVED VOID >1 MM IN NON-TEXTURED, VALVE-TO SHELL-BOND AREA. THE FILL TEST INSPECTION WAS PERFORMED, THE RESULT IS NO BLOCKAGE. BASED ON THE DEVICE ANALYSIS THE FINAL ASSESSMENT IS: A SHARP OPENING ON VALVE BOND EDGE ASSESSED AS AN UNIDENTIFIED (TEAR) OPENING. TREND ANALYSIS: NO PATTERNS WERE FOUND; THEREFORE, NO ADDITIONAL ACTIONS ARE DEEMED REQUIRED. THE TREND OF DEFLATION COMPLAINTS WILL CONTINUE TO BE MONITORED AND CORRECTIVE ACTIONS WILL BE TAKEN IN THE FUTURE IF DEEMED APPROPRIATE.
HEALTHCARE PROFESSIONAL REPORTED LEFT SIDE DEFLATION. DEVICE HAS BEEN EXPLANTED.
A REVIEW OF THE DEVICE HISTORY RECORD HAS BEEN COMPLETED. NO DEVIATIONS OR NON-CONFORMANCES NOTED. FURTHER INFORMATION FROM THE REPORTER REGARDING EVENT, PRODUCT, OR PATIENT DETAILS HAS BEEN REQUESTED. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME. REASON FOR REOPERATION: DEFLATION.
HEALTHCARE PROFESSIONAL REPORTED LEFT SIDE DEFLATION. DEVICE HAS BEEN EXPLANTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 507191 | STYLE 68 SALINE FILLED BREAST IMPLANT | PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE | FWM | ALLERGAN (COSTA RICA) | 2001988 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 63 YR | Required Intervention |