FLO-GARD 6201 VOLUMETRIC INFUSION PUMP
Report
- Report Number
- 6000001-2007-05427
- Event Type
- Malfunction
- Date Received
- September 17, 2008
- Date of Event
- February 1, 2007
- Report Date
- February 5, 2007
- Manufacturer
- BAXTER HEALTHCARE (SG)
- Product Code
- FRN
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER
Narratives
THE FACILITY REPRESENTATIVE REPORTED A CONSTANT OCCLUSION ALARM. INFORMATION WAS NOT PROVIDED REGARDING WHETHER OR NOT THE PROBLEM OCCURRED DURING A PATIENT INFUSION.EVALUATION SUMMARY: THE PUMP WAS EVALUATED BY A BAXTER SERVICE TECHNICIAN. INSPECTION OF THE DEVICE FOUND THAT THE OCCLUSION ALARMS WERE CAUSED BY BROKEN DOOR HINGES AND A BROKEN DOOR LATCH. THE DEVICE WAS RETURNED UNREPAIRED PER CUSTOMER REQUEST. REVIEW OF THE COMPLAINT HISTORY REVEALS SIMILAR REPORTS HAVE BEEN RECEIVED FOR THIS PRODUCT FOR THE REPORTED ISSUE. THIS ISSUE IS BEING INVESTIGATED UNDER CAPA.
THE DEVICE WAS RETURNED FOR SERVICE. DURING SERVICE BY BAXTER, BROKEN DOOR HINGES AND A BROKEN LATCH WERE FOUND. ACCORDING TO THE FACILITY REPRESENTATIVE, NO PATIENT INJURY OR MEDICAL INTERVENTION HAD BEEN REPORTED RELATED TO THE DEVICE. NO ADDITIONAL INFORMATION OR CONTACT WAS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | FLO-GARD 6201 VOLUMETRIC INFUSION PUMP | INFUSION PUMP | FRN | BAXTER HEALTHCARE (SG) | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |