LIFEVEST WCD 3000 SYSTEM
Report
- Report Number
- 3002158293-2008-00432
- Event Type
- Malfunction
- Date Received
- September 8, 2008
- Date of Event
- August 17, 2008
- Report Date
- September 5, 2008
- Manufacturer
- ZOLL LIFECOR CORPORATION
- Product Code
- MVK
- PMA / PMN Number
- P010030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- UNKNOWN
Narratives
DEVICE EVAL SUMMARY: DEVICE EVAL OF ELECTRODE BELT HAS BEEN COMPLETED. THE PROBLEM WAS CONFIRMED. UPON FURTHER INSPECTION, ELECTRODE BELT HAD PINS THAT WERE BENT INSIDE THE CONNECTOR. THE ROOT CAUSE OF THE BENT PINS WAS PROBABLY EXCESSIVE FORCE APPLIED TO THE CONNECTOR DURING MATING. THE CONNECTOR WAS FORCED INTO THE MONITOR WHICH DEFEATED THE CONNECTOR KEYING MECHANISM, AND ALLOWED THE PINS TO MISALIGN WITH THE MATING SOCKETS. THE DAMAGED ELECTRODE BELT CONNECTOR WAS REPLACED. THE ELECTRODE BELT WAS RETESTED AND THEN RESTOCKED. NO ADVERSE EVENT RESULTED FROM THE BENT PINS. THE PT RECEIVED A REPLACEMENT ELECTRODE BELT.
THE HUSBAND OF A FEMALE PT, CONTACTED LIFECOR CUSTOMER SUPPORT TO REPORT THAT THE DEVICE WAS CONTINUALLY STATING TO "ADD GEL/REPLACE BELT". PT'S HUSBAND STATED THAT THERE IS BLUE GEL ON ONLY ONE OF THE THERAPY ELECTRODES. SUPPORT SENT THE PT A REPLACEMENT ELECTRODE BELT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LIFEVEST WCD 3000 SYSTEM | WEARABLE CARDIOVERTER DEFIBRILLATOR | MVK | ZOLL LIFECOR CORPORATION | WCD 3100 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 60 YR |