FDA Adverse Event Malfunction Summary report: N

MWA PERCUTANEOUS ANTENNA 17CM

MDR report key: 1160702 · Received August 29, 2008

Report

Report Number
1717344-2008-00410
Event Type
Malfunction
Date Received
August 29, 2008
Date of Event
August 14, 2008
Report Date
August 18, 2008
Product Code
NEY
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OK, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE IFU STATES: WARNINGS: INSUFFICIENT SPACING BETWEEN ANTENNAS MAY CAUSE EXTENSIVE CHARRING, DIFFICULTY IN REMOVING THE ANTENNA AND INJURY TO THE PT. WHEN USING MULTIPLE MWA ANTENNAS, USE IMAGE GUIDANCE TO ENSURE THAT THE SPACING BETWEEN ANY TWO ANTENNAS IS AT LEAST 1.5 CM.

Description of Event or Problem · 1

THE REPORT STATED THAT FOLLOWING A PERCUTANEOUS MICROWAVE ABLATION, A SKIN BURN WAS FOUND AT THE PENETRATION SITE OF THE MIDDLE ANTENNA OF THREE ANTENNAS USED IN A LINE. THE BURN WAS DESCRIBED AS WHITE WITH A RED RING AROUND THE PENETRATION SITE. THIS WAS A SINGLE ABLATION AT 45W FOR 10 MINS. THE SPACING BETWEEN ANTENNAS WAS ABOUT 1CM TO .5CM IN A LINE. THE TREATMENT WAS A 4X4 MEDICATED ADHESIVE PAD PUT OVER THE PUNCTURE SITES INSTEAD OF THE USUAL BAND-AID. THE ANTENNAS WERE DISCARDED BY THE SITE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MWA PERCUTANEOUS ANTENNA 17CM MICROWAVE ABLATION ACCESSORY NEY UNK

Patients

Seq Age Sex Outcome Treatment
1 UNK