FDA Adverse Event
Malfunction
Summary report: N
ULTRACISION HARMONIC SCALPEL HAND PIECE
MDR report key: 1160675
·
Received September 5, 2008
Report
- Report Number
- 3005075853-2008-01526
- Event Type
- Malfunction
- Date Received
- September 5, 2008
- Date of Event
- August 6, 2008
- Report Date
- August 6, 2008
- Manufacturer
- ETHICON ENDO-SURGERY, LLC
- Product Code
- LFL
- PMA / PMN Number
- K002906
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
DATE SENT: 9/5/2008. EVALUATION SUMMARY THE HAND PIECE WAS RETURNED WITH A CRACKED NOSE CONE AND A LOOSE MOUNT. THE HAND PIECE WAS DISASSEMBLE A TORN ACOUSTIC ISOLATOR AND MOISTURE INGRESS WERE FOUND IN THE INSTRUMENT. THE BATCH HISTORY RECORDS WERE REVIEWED WITH NO ANOMALIES NOTED DURING THE MANUFACTURING PROCESS. COMPLAINT INFORMATION IS TRENDED ON A REGULAR BASIS TO DETERMINE IF FURTHER INVESTIGATION IS WARRANTED.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE ACOUSTIC MOUNT IS LOOSE ON HANDPIECE. NO OTHER INFORMATION AVAILABLE. NO PATIENT CONSEQUENCE REPORTED.