FDA Adverse Event Malfunction Summary report: N

ENDOPATH ETS FLEX45 ENDOSCOPIC ARTICULATING LINEAR CUTTER, 45 MM STAPLE LINE - 2

MDR report key: 1160565 · Received September 11, 2008

Report

Report Number
1527736-2008-03956
Event Type
Malfunction
Date Received
September 11, 2008
Date of Event
June 5, 2008
Report Date
June 9, 2008
Manufacturer
ETHICON ENDO SURGERY, INC (CINCINNATI)
Product Code
GDW
PMA / PMN Number
K020779
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

EVALUATION SUMMARY: THE ANALYSIS RESULTS FOUND THAT THE INSTRUMENT WAS RETURNED WITH THE FIRING MECHANISM DAMAGED AND A FULLY FIRED RELOAD. THE INSTRUMENT WAS DISASSEMBLED TO VERIFY THE CONDITION OF THE INTERNAL COMPONENTS AND THE FIRING TRIGGER TEETH WERE FOUND BROKEN. NO FUNCTIONAL TEST COULD BE PERFORMED DUE TO THE CONDITION OF THE DEVICE. IT SHOULD BE NOTED THAT A 100% INSPECTIONS TAKES PLACE DURING MANUFACTURING TO ENSURE THE DEVICE MEETS THE REQUIRE SPECIFICATIONS; IN ADDITION, A SAMPLE OF THE BATCH IS INSPECTED. THE MANUFACTURING RECORDS WERE REVIEWED AND NO ANOMALIES WERE FOUND DURING THE MANUFACTURING PROCESS.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, THE STAPLES WILL NOT RELEASE. THE DEVICE RELEASED JUST ONE STAPLE (IT WAS NOT POSSIBLE TO OBTAIN MORE INFORMATION). ANOTHER DEVICE WAS USED TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ENDOPATH ETS FLEX45 ENDOSCOPIC ARTICULATING LINEAR CUTTER, 45 MM STAPLE LINE - 2 GDW ETHICON ENDO SURGERY, INC (CINCINNATI) NA E4KEST

Patients

Seq Age Sex Outcome Treatment
1