FDA Adverse Event Malfunction Summary report: N

HUTIV TABLE ASSY STD W/COLOR

MDR report key: 1160563 · Received September 10, 2008

Report

Report Number
1518293-2008-00279
Event Type
Malfunction
Date Received
September 10, 2008
Date of Event
August 25, 2008
Report Date
August 25, 2008
Manufacturer
LIEBEL-FLARSHEIM CO.
Product Code
KQS
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

LIEBEL FLARSHEIM MANUFACTURING REPORT: FIELD SERVICE ENGINEER INVESTIGATED PROBLEM AND FOUND THE TABLE HAD NO 24 VOLT OUTPUT. FSE TRACED PROBLEM TO A J-PLUG CONNECTOR LOOSE ON THE POWER DISTRIBUTION PANEL. RESEATED CONNECTOR AND TIE WRAPPED CABLES TO PREVENT STRESS ON J-PLUG. VERIFIED OPERATION PER HUT IV INSTALLATION AND SERVICE MANUAL. RETURNED TO FULL SERVICE BY THE CUSTOMER.

Description of Event or Problem · 1

CUSTOMER REPORTS VIA E-MAIL. "OVER THE WEEKEND, THE TABLE REFUSED TO MOVE, THERE WAS NO POWER TO THE UNIT." POTENTIAL RISK FOR INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HUTIV TABLE ASSY STD W/COLOR UROLOGY SUITE KQS LIEBEL-FLARSHEIM CO. HUTIV NA

Patients

Seq Age Sex Outcome Treatment
1 UNK