HUTIV TABLE ASSY STD W/COLOR
MDR report key: 1160563
·
Received September 10, 2008
Report
- Report Number
- 1518293-2008-00279
- Event Type
- Malfunction
- Date Received
- September 10, 2008
- Date of Event
- August 25, 2008
- Report Date
- August 25, 2008
- Manufacturer
- LIEBEL-FLARSHEIM CO.
- Product Code
- KQS
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
LIEBEL FLARSHEIM MANUFACTURING REPORT: FIELD SERVICE ENGINEER INVESTIGATED PROBLEM AND FOUND THE TABLE HAD NO 24 VOLT OUTPUT. FSE TRACED PROBLEM TO A J-PLUG CONNECTOR LOOSE ON THE POWER DISTRIBUTION PANEL. RESEATED CONNECTOR AND TIE WRAPPED CABLES TO PREVENT STRESS ON J-PLUG. VERIFIED OPERATION PER HUT IV INSTALLATION AND SERVICE MANUAL. RETURNED TO FULL SERVICE BY THE CUSTOMER.
Description of Event or Problem · 1
CUSTOMER REPORTS VIA E-MAIL. "OVER THE WEEKEND, THE TABLE REFUSED TO MOVE, THERE WAS NO POWER TO THE UNIT." POTENTIAL RISK FOR INJURY.