FDA Adverse Event Malfunction Summary report: N

AFFINITY 4

MDR report key: 1160530 · Received September 10, 2008

Report

Report Number
1824206-2008-03375
Event Type
Malfunction
Date Received
September 10, 2008
Date of Event
August 12, 2008
Report Date
August 13, 2008
Manufacturer
HILL-ROM RITTER
Product Code
HDD
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

HILL-ROM ENGINEERING PERFORMED AN INVESTIGATION ON THE BED AT THE FACILITY, AND FOUND THE MATTRESS HOOK THAT ENGAGES THE LOCKING MECHANISM FOR THE FOOT SECTION (STOW & GO) WAS WORN. ENGINEERING FOUND THAT WHEN THE MATTRESS WAS MOVED FROM SIDE TO SIDE, THE FOOT SECTION LOCKING MECHANISM WOULD UNLOCK. THE FACILITY INFORMED HILL-ROM ENGINEERING THAT A NURSE WAS ALSO INJURED, WHEN WATER WAS SPILLED BY THE PT DURING THE FALL AND THE NURSE SLIPPED AND FELL. THE FACILITY WOULD NOT RELEASE INFO REGARDING THE INJURIES SUSTAINED BY THE PT OR NURSE AND WOULD NOT RELEASE THE PARTS TO HILL-ROM.

Description of Event or Problem · 1

ALLEGED THE PT WAS INJURED BY FALLING BETWEEN THE GAP OF THE SEAT AND FOOT SECTION OF THE BIRTHING BED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 AFFINITY 4 BIRTHING BED HDD HILL-ROM RITTER 3700B

Patients

Seq Age Sex Outcome Treatment
1 UNK Other