FDA Adverse Event Injury Summary report: N

NXSTAGE SYSTEM ONE

MDR report key: 1160466 · Received September 12, 2008

Report

Report Number
3003464075-2008-00424
Event Type
Injury
Date Received
September 12, 2008
Date of Event
August 16, 2008
Report Date
August 16, 2008
Manufacturer
NXSTAGE MEDICAL, INC.
Product Code
KDI
PMA / PMN Number
K050525
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THERE IS NO EVIDENCE OF A DEVICE MALFUNCTION. THE REPORTED BLOOD LOSS EVENT AND VENOUS AIR ALARMS ARE ATTRIBUTED TO USER ERROR AS THE OPERATOR INADVERTENTLY FAILED TO CLAMP THE SALINE LINE AT THE START OF TREATMENT. THE CYCLER ALARMED APPROPRIATELY. THE USER'S GUIDE PROVIDES ADEQUATE INSTRUCTIONS FOR PT CONNECTIONS AND TREATMENT PROCEDURES. A DIRECT CORRELATION BETWEEN NXSTAGE SYSTEM ONE AND THE REPORTED BLOOD LOSS CANNOT BE ESTABLISHED. FACILITY STAFF HAS BEEN NOTIFIED. NXSTAGE MEDICAL CONSIDERS THIS REPORT CLOSED. NO ADDITIONAL INFO WILL BE PROVIDED.

Description of Event or Problem · 1

THIS REPORT IS BEING FILED AS AN ADVERSE EVENT SOLELY TO COMPLY WITH THE FDA'S BLOOD LOSS POLICY. A VENOUS AIR ALARM OCCURRED AT THE START OF A ROUTINE HEMODIALYSIS TREATMENT. THE OPERATOR INADVERTENTLY LEFT THE SALINE LINE UNCLAMPED ALLOWING AIR TO ENTER THE CIRCUIT. RINSEBACK WAS NOT PERFORMED DUE TO THE AMOUNT OF AIR, RESULTING IN AN ESTIMATED BLOOD LOSS OF 190CC. NO MEDICAL INTERVENTION WAS REQUIRED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 NXSTAGE SYSTEM ONE HIGH PERMEABILITY HEMODIALYSIS SYSTEM KDI NXSTAGE MEDICAL, INC. CAR-170 8037703

Patients

Seq Age Sex Outcome Treatment
1 66 YR Other