FDA Adverse Event Malfunction Summary report: N

COLLEAGUE VOLUMETRIC INFUSION PUMP

MDR report key: 1160403 · Received September 16, 2008

Report

Report Number
6000001-2007-03688
Event Type
Malfunction
Date Received
September 16, 2008
Date of Event
January 10, 2007
Report Date
January 10, 2007
Manufacturer
BAXTER HEALTHCARE PTE. LTD.
Product Code
FRN
Removal / Correction Number
CONTINUED ON H10
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VA, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

THIS REPORT IS BEING RESUBMITTED IN ACCORDANCE WITH INSTRUCTIONS FROM FDA TO ADDRESS A MANUFACTURER REPORT SEQUENCE NUMBER ISSUE THAT OCCURRED WHEN THIS MDR OR SUPPLEMENT WAS ORIGINALLY SUBMITTED TO THE FDA ON JAN 31 2007. EVALUATION SUMMARY:EVALUATION WAS PERFORMED AND THE CONDITION WAS CONFIRMED. INSPECTION OF THE PUMP REVEALED DEPLETED MAIN BATTERIES. THE MAIN BATTERIES WERE REPLACED. THE PUMP WAS TESTED FOR PROPER OPERATION AND PUMP FOUND TO FUNCTION PROPERLY. REVIEW OF THE COMPLAINT HISTORY REVEALS SIMILAR REPORTS HAVE BEEN RECEIVED FOR THIS PRODUCT FOR THE REPORTED ISSUE. THIS ISSUE IS CURRENTLY BEING INVESTIGATED UNDER CAPA

Description of Event or Problem · 1

DURING SERVICE BY BAXTER, DEPLETED BATTERIES WERE FOUND. ACCORDING TO THE HOSPITAL REPRESENTATIVE, NO PATIENT INJURY OR MEDICAL INTERVENTION HAD BEEN REPORTED RELATED TO THE DEVICE. NO ADDITIONAL INFORMATION OR CONTACT WAS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COLLEAGUE VOLUMETRIC INFUSION PUMP INFUSION PUMP FRN BAXTER HEALTHCARE PTE. LTD. NA NA

Patients

Seq Age Sex Outcome Treatment
1