FDA Adverse Event Malfunction Summary report: N

COLLEAGUE 3 CX VOLUMETRIC INFUSION PUMP

MDR report key: 1160361 · Received September 16, 2008

Report

Report Number
6000001-2007-03189
Event Type
Malfunction
Date Received
September 16, 2008
Date of Event
January 6, 2007
Report Date
January 6, 2007
Manufacturer
BAXTER HEALTHCARE PTE. LTD.
Product Code
FRN
Removal / Correction Number
CONTINUED FROM H10
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

EVALUATION SUMMARY: THE REPORTED CONDITION OF DEPLETED BATTERIES WAS CONFIRMED. INSPECTION OF THE DEVICE FOUND THAT THE MAIN BATTERIES WERE POTENTIALLY DAMAGED AND WERE THEREFORE REPLACED. REVIEW OF THE COMPLAINT HISTORY REVEALS SIMILAR REPORTS HAVE BEEN RECEIVED FOR THIS PRODUCT FOR THE REPORTED ISSUE. THIS ISSUE IS CURRENTLY BEING INVESTIGATED UNDER CAPA.

Description of Event or Problem · 1

THE FACILITY RETURNED THE DEVICE FOR SERVICE. DURING PRODUCT EVALUATION THE PUMP'S BATTERIES WERE FOUND TO BE DEPLETED. THE HOSPITAL REPRESENTATIVE STATED THAT THERE WERE NO REPORTS OF PATIENT INJURY OR MEDICAL INTERVENTION. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COLLEAGUE 3 CX VOLUMETRIC INFUSION PUMP INFUSION PUMP FRN BAXTER HEALTHCARE PTE. LTD. NA NA

Patients

Seq Age Sex Outcome Treatment
1