ORA SYSTEM 2000
Report
- Report Number
- 2028159-2021-00341
- Event Type
- Malfunction
- Date Received
- April 1, 2021
- Report Date
- July 28, 2021
- Manufacturer
- ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER
- Product Code
- NCF
- PMA / PMN Number
- EXEMPT
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NJ, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
THE SYSTEM WAS EXAMINED. AND THE REPORTED EVENT WAS CONFIRMED AND REPLICATED. THERE WAS A MISALIGNMENT IN THE ABERROMETER, WHICH WAS SUBSEQUENTLY REALIGNED TO RESOLVE THE ISSUE. THE SYSTEM WAS TESTED AND FOUND TO MEET PRODUCT SPECIFICATIONS. THE MANUFACTURING DEVICE HISTORY RECORD (DHR) WAS REVIEWED. BASED ON ASSESSMENT, THE PRODUCT MET SPECIFICATIONS AT THE TIME OF RELEASE. THE ROOT CAUSE OF THE REPORTED EVENT CAN BE ATTRIBUTED TO INTERNAL WEAR/RELIABILITY THAT CAUSED MISALIGNMENT OF THE ABERROMETER. THE MANUFACTURER INTERNAL REFERENCE NUMBER IS: (B)(4).
INVESTIGATION INCLUDING ROOT CAUSE ANALYSIS IS IN PROGRESS. A SUPPLEMENTAL MDR WILL BE FILED AS NECESSARY IN ACCORDANCE WITH 21 CFR 803.56 WHEN ADDITIONAL REPORTABLE INFORMATION BECOMES AVAILABLE. THE MANUFACTURER INTERNAL REFERENCE NUMBER IS: (B)(4).
A HEALTHCARE PROFESSIONAL REPORTED THE SYSTEM HAD INACCURATE READINGS. ADDITIONAL INFORMATION HAS BEEN REQUESTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 498942 | ORA SYSTEM 2000 | ABERROMETER, OPHTHALMIC | NCF | ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER | NA | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | ORA SYSTEM |