FDA Adverse Event
Malfunction
Summary report: N
COLLEAGUE 3CX VOLUMETRIC INFUSION PUMP
MDR report key: 1160331
·
Received September 16, 2008
Report
- Report Number
- 6000001-2007-03662
- Event Type
- Malfunction
- Date Received
- September 16, 2008
- Date of Event
- January 10, 2007
- Report Date
- January 10, 2007
- Manufacturer
- BAXTER HEALTHCARE PTE. LTD.
- Product Code
- FRN
- Removal / Correction Number
- CONTINUED ON H10
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- VA, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
Additional Manufacturer Narrative · 1
EVALUATION SUMMARY: EVALUATION WAS PERFORMED AND THE CONDITION WAS CONFIRMED. INSPECTION OF THE PUMP REVEALED DEPLETED MAIN BATTERIES. THE MAIN BATTERIES WERE REPLACED. THE PUMP WAS TESTED FOR PROPER OPERATION AND THE PUMP FOUND TO FUNCTION PROPERLY. REVIEW OF THE COMPLAINT HISTORY REVEALS SIMILAR REPORTS HAVE BEEN RECEIVED FOR THIS PRODUCT FOR THE REPORTED ISSUE. THIS ISSUE IS CURRENTLY BEING INVESTIGATED UNDER CAPA.
Description of Event or Problem · 1
DURING SERVICE BY BAXTER, DEPLETED BATTERIES WERE FOUND. ACCORDING TO THE HOSPITAL REPRESENTATIVE, NO PATIENT INJURY OR MEDICAL INTERVENTION HAD BEEN REPORTED RELATED TO THE DEVICE. NO ADDITIONAL INFORMATION OR CONTACT WAS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | COLLEAGUE 3CX VOLUMETRIC INFUSION PUMP | INFUSION PUMP | FRN | BAXTER HEALTHCARE PTE. LTD. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |