FDA Adverse Event Malfunction Summary report: N

COLLEAGUE VOLUMETRIC INFUSION PUMP

MDR report key: 1160275 · Received September 16, 2008

Report

Report Number
6000001-2007-00883
Event Type
Malfunction
Date Received
September 16, 2008
Date of Event
December 1, 2006
Report Date
December 12, 2006
Manufacturer
BAXTER HEALTHCARE PTE. LTD.
Product Code
FRN
Removal / Correction Number
CONTINUED ON H10
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

THIS REPORT IS BEING RESUBMITTED IN ACCORDANCE WITH INSTRUCTIONS FROM FDA TO ADDRESS A MANUFACTURER REPORT SEQUENCE NUMBER ISSUE THAT OCCURRED WHEN THIS MDR OR SUPPLEMENT WAS ORIGINALLY SUBMITTED TO THE FDA ON JAN 09 2007. EVALUATION SUMMARY:THE REPORTED CONDITION OF DEPLETED MAIN BATTERIES IDENTIFIED DURING PRODUCT EVALUATION WAS CONFIRMED. INSPECTION OF THE DEVICE FOUND THE PUMP'S MAIN BATTERIES WERE POTENTIALLY DAMAGED AND THEREFORE THE BATTERIES WERE REPLACED. REVIEW OF THE COMPLAINT HISTORY REVEALS SIMILAR REPORTS HAVE BEEN RECEIVED FOR THIS PRODUCT FOR THE REPORTED ISSUE. THIS ISSUE IS CURRENTLY BEING INVESTIGATED. THE EVALUATION OF THE DEVICE WAS PERFORMED ON SITE.

Description of Event or Problem · 1

THE FACILITY REPORTED A PUMP WITH DEPLETED BATTERIES. IT IS UNKNOWN WHEN THIS PROBLEM OCCURRED. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION NECESSARY ACCORDING TO THE HOSPITAL REPRESENTATIVE. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COLLEAGUE VOLUMETRIC INFUSION PUMP INFUSION PUMP FRN BAXTER HEALTHCARE PTE. LTD. NA NA

Patients

Seq Age Sex Outcome Treatment
1