FDA Adverse Event
Malfunction
Summary report: N
COLLEAGUE VOLUMETRIC INFUSION PUMP
MDR report key: 1160265
·
Received September 16, 2008
Report
- Report Number
- 6000001-2007-00580
- Event Type
- Malfunction
- Date Received
- September 16, 2008
- Date of Event
- December 1, 2006
- Report Date
- December 12, 2006
- Manufacturer
- BAXTER HEALTHCARE PTE. LTD.
- Product Code
- FRN
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SC, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THIS REPORT IS BEING RESUBMITTED IN ACCORDANCE WITH INSTRUCTIONS FROM FDA TO ADDRESS A MANUFACTURER REPORT SEQUENCE NUMBER ISSUE THAT OCCURRED WHEN THIS MDR OR SUPPLEMENT WAS ORIGINALLY SUBMITTED TO THE FDA ON JAN 05 2007. EVALUATION SUMMARY:THE INFUSION PUMP WAS TESTED FOR FLOW ACCURACY AT 100 ML/HOUR, THE INFUSION PUMP FAILED THE ACCURACY TEST. THE SPECIFICATION OF THIS DEVICE IS +/-5%. THE PUMP HEAD MODULE WAS REPLACED.
Description of Event or Problem · 1
THE FACILITY REPRESENTATIVE REPORTED AN INFUSION PUMP THAT FAILED THE FLOW RATE TEST DURING BIOMED TESTING THE HOSPITAL REPRESENTATIVE STATED THAT THERE WERE NO REPORTS OF PATIENT INJURY OR MEDICAL INTERVENTION. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | COLLEAGUE VOLUMETRIC INFUSION PUMP | INFUSION PUMP | FRN | BAXTER HEALTHCARE PTE. LTD. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |