COLLEAGUE 3CX VOLUMETRIC INFUSION PUMP
Report
- Report Number
- 6000001-2007-02935
- Event Type
- Malfunction
- Date Received
- September 16, 2008
- Date of Event
- December 1, 2006
- Report Date
- December 12, 2006
- Manufacturer
- BAXTER HEALTHCARE PTE. LTD.
- Product Code
- FRN
- Removal / Correction Number
- 6000001-7/20/05-015-C
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
THIS REPORT IS BEING RESUBMITTED IN ACCORDANCE WITH INSTRUCTIONS FROM FDA TO ADDRESS A MANUFACTURER REPORT SEQUENCE NUMBER ISSUE THAT OCCURRED WHEN THIS MDR OR SUPPLEMENT WAS ORIGINALLY SUBMITTED TO THE FDA ON JAN 25 2007. EVALUATION SUMMARY: THE REPORTED CONDITION OF FAILURE CODE 703 WAS CONFIRMED IN THE PUMP'S EVENT HISTORY FILE DURING PRODUCT EVALUATION. THIS FAILURE CODE WAS CAUSED BY A FAULTY USER INTERFACE MODULE PRINTED CIRCUIT BOARD. THE USER INTERFACE MODULE PRINTED CIRCUIT BOARD WAS REPLACED. REVIEW OF THE COMPLAINT HISTORY REVEALS SIMILAR REPORTS HAVE BEEN RECEIVED FOR THIS PRODUCT FOR THE REPORTED ISSUE. THIS ISSUE IS CURRENTLY BEING INVESTIGATED UNDER CAPA.
THE FIELD SERVICE ENGINEER REPORTED A PUMP WITH FAILURE CODE 703. THIS PROBLEM WAS IDENTIFIED DURING BIO-MED TESTING. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION NECESSARY ACCORDING TO THE HOSPITAL REPRESENTATIVE. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | COLLEAGUE 3CX VOLUMETRIC INFUSION PUMP | INFUSION PUMP | FRN | BAXTER HEALTHCARE PTE. LTD. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |