FDA Adverse Event Malfunction Summary report: N

COLLEAGUE CX VOLUMETRIC INFUSION PUMP

MDR report key: 1160258 · Received September 16, 2008

Report

Report Number
6000001-2007-00321
Event Type
Malfunction
Date Received
September 16, 2008
Date of Event
December 1, 2006
Report Date
December 12, 2006
Manufacturer
BAXTER HEALTHCARE PTE. LTD.
Product Code
FRN
Removal / Correction Number
CONTINUED ON H10
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KS, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

EVALUATION SUMMARY: EVALUATION OCCURRED ON-SITE. THE CONDITION OF 10 BATTERY DISCHARGES BELOW THE ALARM THRESHOLD WAS CONFIRMED IN THE BATTERY HISTORY. THIS INDICATES THAT THE BATTERIES MAY HAVE BEEN DISCHARGED TO A POTENTIALLY DAMAGING LEVEL AND WERE THEREFORE REPLACED. REVIEW OF THE COMPLAINT HISTORY REVEALS SIMILAR REPORTS HAVE BEEN RECEIVED FOR THIS PRODUCT FOR THE REPORTED ISSUE. THIS ISSUE IS BEING INVESTIGATED UNDER CAPA.

Description of Event or Problem · 1

A BAXTER FIELD SERVICE ENGINEER REPORTED AN INFUSION PUMP WITH BATTERIES WITH 10 DISCHARGES BELOW THE ALARM THRESHOLD. INFORMATION WAS NOT PROVIDED REGARDING WHETHER OR NOT THE DEVICE WAS IN PATIENT USE WHEN THE EVENT OCCURRED. ACCORDING TO THE HOSPITAL REPRESENTATIVE, NO PATIENT INJURY OR MEDICAL INTERVENTION HAD BEEN REPORTED RELATED TO THE DEVICE. NO ADDITIONAL INFORMATION OR CONTACT WAS AVAILABLE.