FDA Adverse Event Malfunction Summary report: N

COLLEAGUE 3CX VOLUMETRIC INFUSION PUMP

MDR report key: 1160187 · Received September 16, 2008

Report

Report Number
6000001-2007-01336
Event Type
Malfunction
Date Received
September 16, 2008
Date of Event
December 12, 2006
Report Date
December 12, 2006
Manufacturer
BAXTER HEALTHCARE PTE. LTD.
Product Code
FRN
Removal / Correction Number
CONTINUED ON H10
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AZ, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

EVALUATION SUMMARY: THE CONDITION OF FAILURE CODE 570:320:844:0000 WAS CONFIRMED IN THE PUMP'S EVENT HISTORY FILE DURING PRODUCT EVALUATION. INSPECTION OF THE DEVICE FOUND THAT THE PUMP'S MAIN BATTERIES WERE THIRTY-NINE DISCHARGES BELOW THEIR ALARM THRESHOLD. THE PUMPS MAIN BATTERIES WERE DAMAGED AND THEREFORE THE BATTERIES WERE REPLACED. REVIEW OF THE COMPLAINT HISTORY REVEALS SIMILAR REPORTS HAVE BEEN RECEIVED FOR THIS PRODUCT FOR THE REPORTED ISSUE. THIS ISSUE IS CURRENTLY BEING INVESTIGATED.

Description of Event or Problem · 1

DURING PRODUCT EVALUATION, FAILURE CODE 570:320:844:0000 WAS IDENTIFIED IN THE PUMP'S EVENT HISTORY FILE. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION NECESSARY ACCORDING TO THE HOSPITAL REPRESENTATIVE. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COLLEAGUE 3CX VOLUMETRIC INFUSION PUMP INFUSION PUMP FRN BAXTER HEALTHCARE PTE. LTD. NA NA

Patients

Seq Age Sex Outcome Treatment
1