FDA Adverse Event Malfunction Summary report: N

COLLEAGUE 3CX VOLUMETRIC INFUSION PUMP

MDR report key: 1160132 · Received September 16, 2008

Report

Report Number
6000001-2007-01191
Event Type
Malfunction
Date Received
September 16, 2008
Date of Event
December 12, 2006
Report Date
December 12, 2006
Manufacturer
BAXTER HEALTHCARE PTE. LTD.
Product Code
FRN
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AZ, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THIS REPORT IS BEING RESUBMITTED IN ACCORDANCE WITH INSTRUCTIONS FROM FDA TO ADDRESS A MANUFACTURER REPORT SEQUENCE NUMBER ISSUE THAT OCCURRED WHEN THIS MDR OR SUPPLEMENT WAS ORIGINALLY SUBMITTED TO THE FDA ON JAN 10 2007. EVALUATION SUMMARY: THE CONDITION OF A FAILED ACCURACY TEST FOR UNDERINFUSION ON CHANNEL B WAS CONFIRMED DURING PRODUCT EVALUATION. THIS EVENT WAS CAUSED BY A FAULTY PUMP HEAD MECHANISM ON CHANNEL B. THE PUMP HEAD MECHANISM ON CHANNEL B WAS REPLACED.

Description of Event or Problem · 1

DURING PRODUCT EVALUATION, THE PUMP FAILED A PRE-ACCURACY TEST FOR UNDERINFUSION ON CHANNEL B. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION NECESSARY ACCORDING TO THE HOSPITAL REPRESENTATIVE. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COLLEAGUE 3CX VOLUMETRIC INFUSION PUMP INFUSION PUMP FRN BAXTER HEALTHCARE PTE. LTD. NA NA

Patients

Seq Age Sex Outcome Treatment
1