COLLEAGUE VOLUMETRIC INFUSION PUMP
Report
- Report Number
- 6000001-2007-00293
- Event Type
- Malfunction
- Date Received
- September 16, 2008
- Date of Event
- December 1, 2006
- Report Date
- December 11, 2006
- Manufacturer
- BAXTER HEALTHCARE PTE. LTD.
- Product Code
- FRN
- Removal / Correction Number
- CONTINUED ON H10
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
THIS REPORT IS BEING RESUBMITTED IN ACCORDANCE WITH INSTRUCTIONS FROM FDA TO ADDRESS A MANUFACTURER REPORT SEQUENCE NUMBER ISSUE THAT OCCURRED WHEN THIS MDR OR SUPPLEMENT WAS ORIGINALLY SUBMITTED TO THE FDA ON JAN 03 2007. EVALUATION SUMMARY:EVALUATION OCCURRED ON-SITE. FAILURE CODE 570 WAS CONFIRMED. FAILURE CODE 570 IS CAUSED BY LOW BATTERY VOLTAGE. THE PUMP'S BATTERIES WERE RECHARGED AND THE DEVICE WAS RETURNED TO THE CUSTOMER FULLY OPERATIONAL. REVIEW OF THE COMPLAINT HISTORY REVEALS SIMILAR REPORTS HAVE BEEN RECEIVED FOR THIS PRODUCT FOR THE REPORTED ISSUE. THIS ISSUE IS BEING INVESTIGATED UNDER CAPA.
THE FACILITY REPORTED AN INFUSION PUMP WITH FAILURE CODE 570. THE EVENT WAS REPORTED TO HAVE OCCURRED PRIOR TO PATIENT USE. ACCORDING TO THE HOSPITAL REPRESENTATIVE, NO PATIENT INJURY OR MEDICAL INTERVENTION HAD BEEN REPORTED RELATED TO THE DEVICE. NO ADDITIONAL INFORMATION OR CONTACT WAS AVAILABLE.