FDA Adverse Event Malfunction Summary report: N

IMPL TAPERED SP 3.7MM 12M M HEXAGON

MDR report key: 11601106 · Received April 1, 2021

Report

Report Number
0002023141-2021-00808
Event Type
Malfunction
Date Received
April 1, 2021
Date of Event
February 22, 2021
Report Date
July 15, 2021
Manufacturer
ZIMMER DENTAL
Product Code
DZE
UDI-DI
00889024017474
PMA / PMN Number
K082639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IT
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

ZIMMER BIOMET COMPLAINT NUMBER (B)(4). TWO (2) IMPL TAPERED SP 3.7MM 12M M HEXAGON (SPMB12) WERE RETURNED FOR INVESTIGATION. VISUAL EVALUATION OF THE AS RETURNED PRODUCT IDENTIFIED BOTH IMPLANTS FRACTURED AROUND THE COLLAR REGION. DEVICE HISTORY RECORD (DHR) REVIEW WAS COMPLETED FOR THE SUBJECT LOT NUMBER (2019110202). IT WAS CONFIRMED THAT ALL OPERATIONS AND INSPECTIONS WERE EXECUTED AS PER APPLICABLE PROCEDURE. NO DEVIATIONS OR NON-CONFORMANCES, WHICH COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED EVENT WAS NOTED AS PART OF THE DHR. LOT WAS INSPECTED AND PASSED ALL ACCEPTANCE CRITERIA BY QA. COMPLAINT HISTORY REVIEW WAS PERFORMED FOR THE REPORTED LOT NUMBER (2019110202) FOR SIMILAR EVENT KEYWORD CATEGORY (FUNCTIONAL: FRACTURE : IMPLANT) AND NO OTHER COMPLAINT WAS IDENTIFIED. THEREFORE, BASED ON THE AVAILABLE INFORMATION, DEVICE MALFUNCTION DID OCCUR AND THE REPORTED EVENT WAS CONFIRMED.

Description of Event or Problem · 0

NO FURTHER EVENT INFORMATION IS AVAILABLE AT THE TIME OF THIS REPORT.

Additional Manufacturer Narrative · 1

ZIMMER BIOMET COMPLAINT NUMBER (B)(4). WEIGHT UNKNOWN / NOT PROVIDED. PMA/510(K) NUMBER: K011245 AND K002188.

Description of Event or Problem · 1

IT WAS REPORTED THE IMPLANT FRACTURED AT THE MOMENT OF IMPLANT PLACEMENT. DOCTOR PLACED OTHER IMPLANT IN THE SAME SURGERY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
499870 IMPL TAPERED SP 3.7MM 12M M HEXAGON DENTAL IMPLANT DZE ZIMMER DENTAL SPMB12 2019110202 00889024017474

Patients

Seq Age Sex Outcome Treatment
1 64 YR