FDA Adverse Event Malfunction Summary report: N

FLO-GARD 6301 VOLUMETRIC INFUSION PUMP

MDR report key: 1160074 · Received September 16, 2008

Report

Report Number
6000001-2007-06420
Event Type
Malfunction
Date Received
September 16, 2008
Date of Event
October 31, 2006
Report Date
October 31, 2006
Manufacturer
BAXTER HEALTHCARE
Product Code
FRN
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THIS REPORT IS BEING RESUBMITTED IN ACCORDANCE WITH INSTRUCTIONS FROM FDA TO ADDRESS A MANUFACTURER REPORT SEQUENCE NUMBER ISSUE THAT OCCURRED WHEN THIS MDR OR SUPPLEMENT WAS ORIGINALLY SUBMITTED TO THE FDA ON JUN 07 2007. EVALUATION SUMMARY: THE PUMP WAS EVALUATED BY A BAXTER SERVICE TECHNICIAN. THE REPORTED CONDITION WAS CONFIRMED. REVIEW OF THE COMPLAINT HISTORY REVEALS SIMILAR REPORTS HAVE BEEN RECEIVED FOR THIS PRODUCT FOR THE REPORTED ISSUE. THIS ISSUE IS BEING INVESTIGATED UNDER CAPA.

Description of Event or Problem · 1

THE FACILITY RETURNED THE DEVICE FOR SERVICE. DURING SERVICE TESTING, THE BAXTER REPAIR TECHNICIAN REPORTED A BROKEN DOOR. THE HOSPITAL REPRESENTATIVE STATED THAT THERE HAVE BEEN NO REPORTS OF ANY PATIENT INJURY OR MEDICAL INTERVENTION. NO ADDITIONAL INFORMATION IS KNOWN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 FLO-GARD 6301 VOLUMETRIC INFUSION PUMP INFUSION PUMP FRN BAXTER HEALTHCARE NA NA

Patients

Seq Age Sex Outcome Treatment
1