FDA Adverse Event Malfunction Summary report: N

BABY GORILLA/GORILLA PLATING SYSTEM

MDR report key: 11599922 · Received March 31, 2021

Report

Report Number
3008650117-2021-00075
Event Type
Malfunction
Date Received
March 31, 2021
Date of Event
June 15, 2020
Report Date
March 31, 2021
Manufacturer
PARAGON 28, INC.
Product Code
HTJ
PMA / PMN Number
K190365
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE HISTORY RECORDS FOR LOT NUMBER 9151337 WERE REVIEWED AND SHOWS ALL INSPECTED PARTS WERE RECEIVED AND ACCEPTED. VISUAL OBSERVATION OF THE INSTRUMENT CONFIRMED THE DEPTH GAUGE BROKEN INTO PIECES. IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL REPORT WILL BE FILED ACCORDINGLY.

Description of Event or Problem · 1

A PARGON 28 DEPTH GAUGE PLATE SCREW, 70MM SCREW WAS REPORTED BROKEN AFTER BEING WASHED. THE NEEDLE OF THE DEVICE IS SNAPPED OFF COMPLETELY. IT WAS REPORTED THAT THE INSTRUMENT WAS NOT BROKEN IN SURGERY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
497872 BABY GORILLA/GORILLA PLATING SYSTEM DEPTH GAUGE PLATE SCREW, 70MM SCREW HTJ PARAGON 28, INC. P99-150-0014 9151337

Patients

Seq Age Sex Outcome Treatment
1