Summary report: N
COLLEAGUE CX VOLUMETRIC INFUSION PUMP
MDR report key: 1159830
·
Received September 16, 2008
Report
- Report Number
- 6000001-2007-01515
- Event Type
- Malfunction
- Date Received
- September 16, 2008
- Date of Event
- October 18, 2006
- Report Date
- October 18, 2006
- Manufacturer
- BAXTER HEALTHCARE PTE. LTD.
- Product Code
- FRN
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AZ, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
Additional Manufacturer Narrative · 1
THIS REPORT IS BEING RESUBMITTED IN ACCORDANCE WITH INSTRUCTIONS FROM FDA TO ADDRESS A MANUFACTURER REPORT SEQUENCE NUMBER ISSUE THAT OCCURRED WHEN THIS MDR OR SUPPLEMENT WAS ORIGINALLY SUBMITTED TO THE FDA ON JAN 16 2007. EVALUATION SUMMARY:THE CONDITION OF INACCURACY WAS CONFIRMED. INSPECTION OF THE DEVICE FOUND THAT THIS WAS CAUSED BY A FAULTY PUMP HEAD MODULE THAT WAS REPLACED.
Description of Event or Problem · 1
THE FACILITY RETURNED THE DEVICE FOR SERVICE. DURING SERVICE TESTING, THE BAXTER REPAIR TECHNICIAN REPORTED THAT THE DEVICE WAS INACCURATE.