INSPIRA TEXTURED SILICONE GEL FILLED BREAST IMPLANT
Report
- Report Number
- 9617229-2021-04071
- Event Type
- Injury
- Date Received
- March 30, 2021
- Date of Event
- November 16, 2020
- Report Date
- March 30, 2021
- Manufacturer
- ALLERGAN (COSTA RICA)
- Product Code
- FTR
- PMA / PMN Number
- P020056
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SP
- Reporter Occupation
- PHYSICIAN
Narratives
DEVICE EVALUATION: VISUAL ANALYSIS OF THE RETURNED DEVICE IDENTIFIED ONE EXTENDED OPENING. A WEIGHT TEST OF THE DEVICE WAS VERIFIED AND THE DEVICE UNDERWEIGHT. A MICROSCOPIC ANALYSIS WAS PERFORMED WHICH IDENTIFY: ONE SHARP EDGE OPENING IN THE SHELL WITH STRESS MARKS. A DIMENSION MEASUREMENT IN THE SHELL WAS PERFORMED WHICH IDENTIFY THE THICKNESS WITHIN SPECIFICATION. BASED ON THE DEVICE ANALYSIS THE FINAL ASSESSMENT IS: A SHARP EDGE OPENING IN THE SHELL WITH STRESS MARKS IN THE SIDE POSTERIOR ASSESSED AS SURGICAL IMPACT OPENING.
HEALTHCARE PROFESSIONAL REPORTED RIGHT SIDE RUPTURE. DEVICE HAS BEEN EXPLANTED.
A REVIEW OF THE DEVICE HISTORY RECORD HAS BEEN COMPLETED. NO DEVIATIONS OR NON-CONFORMANCES NOTED. DEVICE PHOTO EVALUATION: VISUAL ANALYSIS OF THE PHOTOGRAPHS IDENTIFIED ONE EXTENDED OPENING AND RED ENCAPSULATION. FURTHER INFORMATION FROM THE REPORTER REGARDING EVENT, PRODUCT, OR PATIENT DETAILS HAS BEEN REQUESTED. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME. REASON FOR REOPERATION: RUPTURE.
HEALTHCARE PROFESSIONAL REPORTED RIGHT SIDE RUPTURE. DEVICE HAS BEEN EXPLANTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 484611 | INSPIRA TEXTURED SILICONE GEL FILLED BREAST IMPLANT | PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED | FTR | ALLERGAN (COSTA RICA) | 2683734 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 44 YR | Required Intervention |