FDA Adverse Event Injury Summary report: N

INSPIRA TEXTURED SILICONE GEL FILLED BREAST IMPLANT

MDR report key: 11588669 · Received March 30, 2021

Report

Report Number
9617229-2021-04071
Event Type
Injury
Date Received
March 30, 2021
Date of Event
November 16, 2020
Report Date
March 30, 2021
Manufacturer
ALLERGAN (COSTA RICA)
Product Code
FTR
PMA / PMN Number
P020056
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
SP
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 0

DEVICE EVALUATION: VISUAL ANALYSIS OF THE RETURNED DEVICE IDENTIFIED ONE EXTENDED OPENING. A WEIGHT TEST OF THE DEVICE WAS VERIFIED AND THE DEVICE UNDERWEIGHT. A MICROSCOPIC ANALYSIS WAS PERFORMED WHICH IDENTIFY: ONE SHARP EDGE OPENING IN THE SHELL WITH STRESS MARKS. A DIMENSION MEASUREMENT IN THE SHELL WAS PERFORMED WHICH IDENTIFY THE THICKNESS WITHIN SPECIFICATION. BASED ON THE DEVICE ANALYSIS THE FINAL ASSESSMENT IS: A SHARP EDGE OPENING IN THE SHELL WITH STRESS MARKS IN THE SIDE POSTERIOR ASSESSED AS SURGICAL IMPACT OPENING.

Description of Event or Problem · 0

HEALTHCARE PROFESSIONAL REPORTED RIGHT SIDE RUPTURE. DEVICE HAS BEEN EXPLANTED.

Additional Manufacturer Narrative · 1

A REVIEW OF THE DEVICE HISTORY RECORD HAS BEEN COMPLETED. NO DEVIATIONS OR NON-CONFORMANCES NOTED. DEVICE PHOTO EVALUATION: VISUAL ANALYSIS OF THE PHOTOGRAPHS IDENTIFIED ONE EXTENDED OPENING AND RED ENCAPSULATION. FURTHER INFORMATION FROM THE REPORTER REGARDING EVENT, PRODUCT, OR PATIENT DETAILS HAS BEEN REQUESTED. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME. REASON FOR REOPERATION: RUPTURE.

Description of Event or Problem · 1

HEALTHCARE PROFESSIONAL REPORTED RIGHT SIDE RUPTURE. DEVICE HAS BEEN EXPLANTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
484611 INSPIRA TEXTURED SILICONE GEL FILLED BREAST IMPLANT PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED FTR ALLERGAN (COSTA RICA) 2683734

Patients

Seq Age Sex Outcome Treatment
1 44 YR Required Intervention