FDA Adverse Event Injury Summary report: N

FLUOROSTAR

MDR report key: 11586514 · Received March 29, 2021

Report

Report Number
9680959-2021-00001
Event Type
Injury
Date Received
March 29, 2021
Date of Event
January 23, 2021
Report Date
September 8, 2021
Manufacturer
GE OEC MEDICAL SYSTEMS GMBH 14 WILHELM-MAISEL-STR
Product Code
OWB
PMA / PMN Number
K043076
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
BR
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

GE HEALTHCARE'S INVESTIGATION INTO THE REPORTED EVENT HAS CONCLUDED THAT AFTER ADDITIONAL INFORMATION WAS RECEIVED FROM THE CUSTOMER IT WAS CONFIRMED THAT A SERIOUS INJURY DID NOT OCCUR. A REVIEW OF THE EVENT BY THE GE OEC MEDICAL DIRECTOR CONCLUDED THAT, SHOULD THE EVENT RECUR, IT WOULD NOT BE LIKELY TO RESULT IN DEATH OR SERIOUS INJURY. THEREFORE THIS EVENT IS NOT REPORTABLE UNDER 21 CFR 803.

Additional Manufacturer Narrative · 1

THE CUSTOMER REPORTED WHILE PERFORMING A PERCUTANEOUS NEPHROLITHOTOMY, DURING LIVE IMAGE ACQUISITION, THE IMAGE BECAME DARK, THE SYSTEM LOST FUNCTIONALITY AND AN ERROR MESSAGE WAS DISPLAYED PREVENTING THE CONTINUATION OF THE SURGERY. THE CLINICAL TEAM CHOSE TO TERMINATE THE PROCEDURE RESULTING IN PROLONGATION OF HOSPITALIZATION TIME. THE PROCEDURE WAS SUCCESSFULLY COMPLETED ON A LATER DATE AND THE PATIENT IS RECOVERING AT HOME WITHOUT COMPLICATIONS. GE HEALTHCARES INVESTIGATION INTO THE REPORTED OCCURRENCE IS ONGOING. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE INVESTIGATION HAS BEEN COMPLETED, IF ADDITIONAL INFORMATION IS IDENTIFIED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED WHILE PERFORMING A PERCUTANEOUS NEPHROLITHOTOMY, DURING LIVE IMAGE ACQUISITION, THE IMAGE BECAME DARK, THE SYSTEM LOST FUNCTIONALITY AND AN ERROR MESSAGE WAS DISPLAYED PREVENTING THE CONTINUATION OF THE SURGERY. THE CLINICAL TEAM CHOSE TO TERMINATE THE PROCEDURE RESULTING IN PROLONGATION OF HOSPITALIZATION TIME. THE PROCEDURE WAS SUCCESSFULLY COMPLETED ON A LATER DATE AND THE PATIENT IS RECOVERING AT HOME WITHOUT COMPLICATIONS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
482430 FLUOROSTAR FLUOROSCOPIC X-RAY OWB GE OEC MEDICAL SYSTEMS GMBH 14 WILHELM-MAISEL-STR XMS577 79-C5208D

Patients

Seq Age Sex Outcome Treatment
1 51 YR Hospitalization