ID NOW COVID-19 ASSAY
Report
- Report Number
- 1221359-2021-00476
- Event Type
- Malfunction
- Date Received
- March 27, 2021
- Date of Event
- May 18, 2020
- Report Date
- April 24, 2021
- Manufacturer
- ABBOTT DIAGNOSTICS SCARBOROUGH, INC.
- Product Code
- QJR
- UDI-DI
- 10811877011269
- PMA / PMN Number
- EUA2000074
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MT, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
ADDITIONAL INFORMATION/CORRECTED DATA: ABBOTT DIAGNOSTICS SCARBOROUGH WAS NOT ABLE TO ABLE TO DETERMINE THE EXACT ROOT CAUSE OF THE POSITIVE RESULT; HOWEVER IT COULD POSSIBLY BE RELATED TO THE SPECIFIC PATIENT SAMPLE.
TESTING WAS PERFORMED AT ABBOTT DIAGNOSTICS (B)(4), INC. ON RETAINED KIT LOT M119307 WITH INTERNAL POSITIVE QUALITY CONTROL SAMPLES AND NEGATIVE QUALITY CONTROL SWABS. ALL TEST RESULTS WERE VALID AND PERFORMED AS EXPECTED. ADDITIONALLY, THE MANUFACTURING RECORDS AND QUALITY CONTROL RELEASE TESTING WAS REVIEWED FOR TEST KIT PART NUMBER 190-000 / LOT M119307, TEST BASE PART NUMBER 190-430 / LOT M119307. THE LOT MET THE REQUIRED RELEASE SPECIFICATIONS. A REVIEW OF THE COMPLAINTS REPORTED AS FALSE POSITIVE PATIENT RESULTS (CONFIRMED AND UNCONFIRMED, CONFLICTING RESULTS) RELATED TO KIT LOT M119307 SHOWED THAT THE COMPLAINT RATE IS (B)(4). ABBOTT DIAGNOSTICS (B)(4), INC. WAS UNABLE TO DETERMINE THE EXACT ROOT CAUSE OF THE REPORTED ISSUE. THE AVAILABLE EVIDENCE SUGGESTS THAT THIS DEVICE LOT IS PERFORMING WITHIN THE STATEMENTS MADE WITHIN THE PACKAGE INSERT RELATED TO % TEST AGREEMENT WITH THE EXPECTED RESULTS BY SAMPLE CONCENTRATION.
THE CUSTOMER REPORTED A FALSE POSITIVE RESULT WITH THE ID NOW COVID-19 ASSAY PERFORMED ON (B)(6) 2020 ON A DIRECT TESTED NASAL KITTED SWAB OR COPAN NP SWAB. ON (B)(6) 2020 REPEAT TESTING WAS PERFORMED WITH A DIRECT TESTED NASAL KITTED SWAB OR COPAN NASOPHARYNGEAL SWAB WITH A NEGATIVE RESULT. ON (B)(6) 2020, CONFIRMATION TESTING WITH A NASOPHARYNGEAL SWAB IN VTM WAS PERFORMED ON THERMOFISHER ASSAY, KINGFISHER FOR EXTRACTION; ABI 7500 FAST DX FOR AMPICATION AND DETECTION, NEGATIVE RESULT WITH ONLY CT VALUE OF THE CONTROL, 27.9, PROVIDED. THE CUSTOMER STATED THE PATIENT WAS ASYMPTOMATIC. THE CUSTOMER CONFIRMED THERE WAS NO PATIENT HARM DUE TO THE TEST RESULTS ALTHOUGH AN ELECTIVE OTHOPEDIC SURGERY WAS RESCHEDULED DUE TO THE TEST RESULTS. POSITIVE RESULTS ARE INDICATIVE OF THE PRESENCE OF SARS-COV-2 RNA; CLINICAL CORRELATION WITH PATIENT HISTORY AND OTHER DIAGNOSTIC INFORMATION IS NECESSARY TO DETERMINE PATIENT INFECTION STATUS. POSITIVE RESULTS DO NOT RULE OUT BACTERIAL INFECTION OR CO-INFECTION WITH OTHER VIRUSES. DUE TO THE HIGH SENSITIVITY OF THE ASSAYS RUN ON THE INSTRUMENT, CONTAMINATION OF THE WORK AREA WITH PREVIOUS POSITIVE SAMPLES MAY CAUSE FALSE POSITIVE RESULTS. HANDLE SAMPLES ACCORDING TO STANDARD LABORATORY PRACTICES. CLEAN INSTRUMENTS AND SURROUNDING SURFACES ACCORDING TO INSTRUCTIONS PROVIDED IN THE CLEANING SECTION OF THE INSTRUMENT USER MANUAL. DUE TO THE RISK OF A FALSE POSITIVE RESULT LEADING TO POSSIBLE EXPOSURE TO COVID-19 THROUGH THE QUARANTINE OF THE FALSE POSITIVE PATIENT WITH TRUE POSITIVE PATIENT, THE INCIDENT SHALL BE CONSIDERED REPORTABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 476128 | ID NOW COVID-19 ASSAY | MOLECULAR IVD FOR ID NOW COVID-19 | QJR | ABBOTT DIAGNOSTICS SCARBOROUGH, INC. | M119307 | 10811877011269 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 17 YR |