EQUINOXE
Report
- Report Number
- 1038671-2021-00132
- Event Type
- Injury
- Date Received
- March 26, 2021
- Date of Event
- March 2, 2021
- Report Date
- May 12, 2021
- Manufacturer
- EXACTECH, INC.
- Product Code
- KWT
- UDI-DI
- 10885862086419
- PMA / PMN Number
- K110708
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NH, US
- Reporter Occupation
- PHYSICIAN
Narratives
SECTION H10: (H3) THE REVISION REPORTED WAS LIKELY THE RESULT OF POOR BONE QUALITY AND AN INSUFFICIENT BOND BETWEEN THE GLENOID BASEPLATE AND THE BONE, WHICH LED TO ASEPTIC (NON-INFECTED) GLENOID LOOSENING. HOWEVER, THIS CANNOT BE CONFIRMED AS THE DEVICES WERE NOT AVAILABLE FOR EVALUATION. SECTION H11: *THE FOLLOWING SECTIONS HAVE CORRECTED INFORMATION: (D2) COMMON DEVICE NAME: RS GLENOID PLATE R POST AUG, 8 DEG, RIGHT (D4) CATALOG NUMBER: 320-15-04, SERIAL NUMBER: (B)(6) EXPIRATION DATE: 21-APR-2029, UNIQUE IDENTIFIER (UDI) #: (B)(4). (G4) PMA/510(K)NUMBER: K110708 (H4) DEVICE MANUFACTURE DATE: 22-APR-2019 (D10) CONCOMITANT DEVICES: 320-10-00, 6159881 - EQUINOXE REVERSE TRAY ADAPTER PLATE TRAY +0 320-38-00, 6054852 - EQUINOXE REVERSE 38MM HUMERAL LINER +0 320-02-38, 5372918 - RS EXP GLENOSPHERE 38M M+4 320-20-00, 6136875 - RS TORQUE SCREW KIT 320-20-22, 4519568 - 22MM COMPRESSION SCREW 320-15-05, 6152600 - EQ REV LOCKING SCREW 320-20-22, 5494681 - 22MM COMPRESSION SCREW 320-20-26, 5901911 - 26MM COMPRESSION SCREW 320-20-26, 6104386 - 26MM COMPRESSION SCREW 320-20-30, 5961269 - 30MM COMPRESSION SCREW TPA-13, AB4152 - XS CEMENT RESTRICTOR
CONCOMITANT DEVICE(S): 320-02-38, 5372918 - RS EXP GLENOSPHERE 38M M+4. 320-10-00, 6159881 - RS ADAPTER TRAY. 320-15-04, 5976414 - RS GLENOID PLT 8 DEG RIGHT POST AUG. 320-20-00, 6136875 - RS TORQUE SCREW KIT. 320-20-22, 4519568 - 22MM COMPRESSION SCREW. 320-15-05, 6152600 - EQ REV LOCKING SCREW. 320-20-22, 5494681 - 22MM COMPRESSION SCREW. 320-20-26, 5901911 - 26MM COMPRESSION SCREW. 320-20-26, 6104386 - 26MM COMPRESSION SCREW. 320-20-30, 5961269 - 30MM COMPRESSION SCREW. TPA-13, AB4152 - XS CEMENT RESTRICTOR.
AS REPORTED, APPROXIMATELY 26 MONTHS AFTER THE INITIAL R TSA, THE (B)(6) Y/O FEMALE PATIENT WAS EXPERIENCING PAIN IN THE REVISED RIGHT REVERSE SHOULDER. HER BONE QUALITY WAS DETERMINED TO BE VERY BAD AT A PRIOR REVISION. X-RAYS SHOWED A LOOSE BASEPLATE. GIVEN HEALTH OF PATIENT, SURGEON CHOSE TO REPLACE THE REVERSE SHOULDER FOR A "BIG BALL HEMI" TO RELIEVE PAIN AND ALLOW FOR SOME FUNCTION. EXPLANTS WILL NOT BE RETURNED PER HOSPITAL. PATIENT IS EXPECTED TO RECOVER WELL AND LEFT THE OR STABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 471293 | EQUINOXE | RS GLENOID PLATE R POST AUG, 8 DEG, RIGHT | KWT | EXACTECH, INC. | 320-10-00 | UNK | 10885862086419 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 79 YR | Hospitalization| R |