REDUCED SIZE SCREW EXPANDABLE LEFT DISTAL FEMORAL 20CM REPLACEMENT
Report
- Report Number
- 0001825034-2021-00904
- Event Type
- Injury
- Date Received
- March 26, 2021
- Date of Event
- February 17, 2021
- Report Date
- July 9, 2021
- Manufacturer
- ZIMMER BIOMET, INC.
- Product Code
- JWH
- PMA / PMN Number
- CUSTOM
- Removal / Correction Number
- N/A
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IN, US
- Reporter Occupation
- PHYSICIAN
Narratives
COMPLAINT SAMPLE WAS EVALUATED. VISUAL EVALUATION OF THE RETURNED DEVICE SHOWS THAT COMPONENTS CP-11-1829-01 AND CP-11-1343-03 TO BE FRACTURED AND SIGNS OF BEING IMPLANTED FOR 7 YEARS, WEAR AND BIOMATERIAL. DEVICE HISTORY RECORD (DHR) WAS REVIEWED AND NO DISCREPANCIES RELEVANT TO THE REPORTED EVENT WERE FOUND. A DEFINITIVE ROOT CAUSE CANNOT BE DETERMINED. IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL WILL BE FILED ACCORDINGLY. ZIMMER BIOMET WILL CONTINUE TO MONITOR FOR TRENDS.
NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.
(B)(4). CONCOMITANT MEDICAL DEVICES - ORTHOPEDIC SALVAGE SYSTEM CPS TRANSVERSE PIN 32MM CATALOG #: 178527 LOT #: 280870, ORTHOPEDIC SALVAGE SYSTEM CPS ANCHOR PLUG 12MM CATALOG #: 178402 LOT #: 727330, ORTHOPEDIC SALVAGE SYSTEM CPS SPINDLE WITH PINS 600LBF CATALOG #: 178354 LOT #: 349010, ORTHOPEDIC SALVAGE SYSTEM CPS NUT CATALOG #: 178512 LOT #: 599470, ORTHOPEDIC SALVAGE SYSTEM RS POLY TIBIAL BEARING 16 CATALOG #: 161030 LOT #: 040940, ORTHOPEDIC SALVAGE SYSTEM RS NON-MODULAR PLATE LONG 71 CATALOG #: 161041 LOT #: 239710, ORTHOPEDIC SALVAGE SYSTEM RS POLY FEMORAL BUSHINGS CATALOG #: 161034 LOT #: 833940, ORTHOPEDIC SALVAGE SYSTEM POLY LOCKING PIN CATALOG #: 150478 LOT #: 716500, ORTHOPEDIC SALVAGE SYSTEM RS AXLE CATALOG #: 161035 LOT #: 859600, ORTHOPEDIC SALVAGE SYSTEM REINFORCED YOKE CATALOG #: 150493 LOT #: 194890, ORTHOPEDIC SALVAGE SYSTEM POLY TIBIAL BUSHING CATALOG #: 150476 LOT #: 901590, SERIES A STANDARD 3 PEG PATELLA 31MM CATALOG #: 184764 LOT #: 013860. THE PRODUCT HAS BEEN RECEIVED BY ZIMMER BIOMET AND THE INVESTIGATION IS IN PROCESS. ONCE THE INVESTIGATION HAS BEEN COMPLETED, A FOLLOW-UP MDR WILL BE SUBMITTED.
IT WAS REPORTED THAT THE PATIENT UNDERWENT A KNEE ARTHROPLASTY REVISION TO ADDRESS INSTABILITY AND BREAKAGE OF THE CUSTOM EXPANDABLE IMPLANT APPROXIMATELY SEVEN (7) YEARS POST-OPERATIVELY. ATTEMPTS HAVE BEEN MADE, HOWEVER, NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 470315 | REDUCED SIZE SCREW EXPANDABLE LEFT DISTAL FEMORAL 20CM REPLACEMENT | PROSTHESIS, KNEE | JWH | ZIMMER BIOMET, INC. | N/A | 993290 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Hospitalization| R | SEE H10 NARRATIVE |