FDA Adverse Event Injury Summary report: N

STEM: QUADRA-H CEMENTLESS, HA COATED LAT STEM SIZE 1

MDR report key: 11568315 · Received March 25, 2021

Report

Report Number
3005180920-2021-00235
Event Type
Injury
Date Received
March 25, 2021
Date of Event
February 26, 2021
Report Date
March 25, 2021
Manufacturer
MEDACTA INTERNATIONAL SA
Product Code
JDI
UDI-DI
07630030802133
PMA / PMN Number
K082792
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 1

BATCH REVIEW PERFORMED ON 04 MARCH 2021: LOT 072273: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 18-OCT-2007. EXPIRATION DATE: 2012-09-31. NO ANOMALIES FOUND RELATED TO THE PROBLEM. (B)(4) ITEMS HAS BEEN RE-STERILIZED AS LOT 072273/T AND RELEASED ON 07-NOV-2013. EXPIRATION DATE: 2018-09-31. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD WITHOUT ANY SIMILAR REPORTED EVENT SINCE 1-1-2017. CLINICAL EVALUATION: STEM REVISION PERFORMED 4 YEARS AND 8 MONTHS AFTER PRIMARY CEMENTLESS TOTAL HIP ARTHROPLASTY IN A (B)(6) YEAR OLD MAN. IN THE RADIOGRAPHIC IMAGE PROVIDED, SIGNS OF STRESS SHIELDING ARE VISIBLE AND THE STEM LOOKS SLIGHTLY UNDERSIZED. THE REASON OF THIS CHOICE CANNOT BE ASSESSED ON THE BASIS OF A SINGLE ANTEROPOSTERIOR RADIOGRAPHIC PROJECTION. ASEPTIC LOOSENING IS A POSSIBLE LITERATURE DESCRIBED ADVERSE EVENT AFTER PRIMARY CEMENTLESS HIP ARTHROPLASTIES AND CAUSES ARE OFTEN UNKNOWN. THE REASON OF THIS FAILURE CANNOT BE DETERMINED. PRELIMINARY INVESTIGATION: LOOKING AT THE IMAGES OF THE EXPLANTED STEM IT IS VISIBLE THAT ALMOST ALL HA COATING ON THE STEM BODY HAS BEEN ABSORBED BY PATIENT BONE AS EXPECTED. THERE ARE ONLY A FEW RESIDUAL FRAGMENTS OF HA COATING ON THE BODY. THE ROOT CAUSE REMAINS UNKNOWN.

Description of Event or Problem · 1

4 YEARS AND 8 MONTHS AFTER PRIMARY, REVISION SURGERY FOR LOOSE QUADRA H STEM. THE SURGEON REVISED SUCCESSFULLY THE STEM AND HEAD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
464871 STEM: QUADRA-H CEMENTLESS, HA COATED LAT STEM SIZE 1 CEMENTLESS STEM JDI MEDACTA INTERNATIONAL SA 01.12.031 072273/T 07630030802133

Patients

Seq Age Sex Outcome Treatment
1 71 YR Required Intervention