FDA Adverse Event Malfunction Summary report: N

SOCKET WRENCH FOR VEPTR NUT

MDR report key: 11567994 · Received March 25, 2021

Report

Report Number
2939274-2021-01515
Event Type
Malfunction
Date Received
March 25, 2021
Report Date
February 25, 2021
Manufacturer
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
Product Code
HXC
UDI-DI
10705034720128
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CO, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

ADDITIONAL NARRATIVE: REPORTER IS A SYNTHES EMPLOYEE. PART NUMBER:03.641.004. SYNTHES LOT NUMBER: H852117-11. SUPPLIER LOT NUMBER: N/A. RELEASE TO WAREHOUSE DATE: (B)(4) 2020. EXPIRATION DATE: N/A. SUPPLIER: AVALIGN TECHNOLOGIES-NEMCOMED. NO NCRS WERE GENERATED DURING PRODUCTION. REVIEW OF THE DEVICE HISTORY RECORD SHOWED THAT THERE WERE NO ISSUES DURING THE MANUFACTURE OF THIS PRODUCT, AND ANY SUBCOMPONENTS, WHICH WOULD CONTRIBUTE TO THIS COMPLAINT CONDITION. SERVICE AND REPAIR EVALUATION: DURING TESTING AT SERVICE AND REPAIR IT WAS FOUND OUT THAT THE SOCKET WRENCH HAD FAILED IN CALIBRATION. THE REPAIR TECHNICIAN REPORTED THE DEVICE FAILED TORQUE TESTING. THE CAUSE OF THE ISSUE IS FAILED HIGH. THE ITEM WILL BE REPAIRED AND WILL BE RETURNED TO THE CUSTOMER UPON COMPLETION OF THE SERVICE AND REPAIR PROCESS. THE EVALUATION WAS CONFIRMED. THE DEVICE WAS DEEMED SERVICEABLE AND WILL BE RETURNED TO THE CUSTOMER, NO DESIGN OR MANUFACTURING ISSUES WERE IDENTIFIED THEREFORE IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Description of Event or Problem · 1

IT WAS REPORTED THAT ON (B)(4) 2021, DURING EVALUATION AT SERVICE AND REPAIR THE SOCKET WRENCH FAILED IN CALIBRATION. THERE WAS NO PATIENT INVOLVEMENT. THIS REPORT INVOLVES ONE (1) SOCKET WRENCH FOR VEPTR NUT. THIS IS REPORT 1 OF 1 FOR PC (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
465813 SOCKET WRENCH FOR VEPTR NUT WRENCH HXC WRIGHTS LANE SYNTHES USA PRODUCTS LLC 03.641.004 H852117-11 10705034720128

Patients

Seq Age Sex Outcome Treatment
1