FDA Adverse Event Injury Summary report: N

MENTOR SILTEX ROUND MODERATE PROFILE

MDR report key: 11566085 · Received March 24, 2021

Report

Report Number
1645337-2021-03132
Event Type
Injury
Date Received
March 24, 2021
Date of Event
February 6, 2021
Manufacturer
MENTOR TEXAS
Product Code
FWM
UDI-DI
00081317001829
PMA / PMN Number
P990075
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MO, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

THE MENTOR FAILURE ANALYSIS LAB RECEIVED THE DEVICE FOR EVALUATION ON (B)(6) 2021. THE ANALYSIS HAS BEGUN BUT IS NOT COMPLETE AT THIS TIME. WHEN THE INVESTIGATIONAL ANALYSIS HAS BEEN COMPLETED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. REPLACEMENT WITH A 425CC MENTOR SMOOTH ROUND MODERATE PROFILE SALINE PROSTHESIS WAS PERFORMED WITH EXPLANT. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).

Additional Manufacturer Narrative · 0

THE INVESTIGATION OF THE RETURNED DEVICE WAS COMPLETED BY THE FAILURE ANALYSIS LAB ON JUNE 24, 2021. DEVICE EVALUATION SUMMARY: THE PRODUCT WAS RETURNED TO MENTOR FOR EVALUATION. MENTOR CONDUCTED A VISUAL INSPECTION OF THE RETURNED DEVICE. VISUAL ANALYSIS OF THE RETURNED SAMPLE REVEALED THAT THE BREAST IMPLANT HAD AN AREA OF SILTEX CRACKING ON THE POSTERIOR VIEW. ADDITIONALLY, A TEAR MEASURING APPROXIMATELY 0.6 CM WAS FOUND WITHIN THE SILTEX CRACKING. THE EVALUATION DETERMINED THAT THE CAUSE OF THE DEFLATION IS CONSISTENT WITH NORMAL WEAR. SILTEX CRACKING IS DEFINED AS A MATERIAL SEPARATION OCCURRING IN THE SILTEX LAYER AND CAN EVENTUALLY PROGRESS THROUGH THE SHELL OF SILTEX DEVICES. SILTEX CRACKING IS ULTIMATELY A RESULT OF HIGH STRESSES. IT SHOULD BE NOTED THAT AS PART OF MENTOR¿S QUALITY PROCESS ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED TO APPROVED SPECIFICATIONS. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).

Additional Manufacturer Narrative · 0

A MANUFACTURING RECORD EVALUATION WAS PERFORMED FOR THE FINISHED DEVICE 220652 NUMBER, AND NO NON-CONFORMANCES WERE IDENTIFIED. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).

Additional Manufacturer Narrative · 0

ADDITIONAL INFORMATION WAS RECEIVED ON (B)(6) 2021. THE DEVICE WAS EXPLANTED ON (B)(6) 2021. 2. IS OTHER SERIOUS- UNCHECK. 2. IS REQUIRED INTERVENTION- CHECK . SINCE THE DEVICE HAS NOT BEEN RETURNED FOR ANALYSIS, NO PRODUCT FAILURE ANALYSIS CAN BE CONDUCTED, AND NO DETERMINATION OF POSSIBLE CONTRIBUTING FACTORS CAN BE MADE. AS SUCH, THE INVESTIGATION WILL BE CLOSED. IF THE COMPLAINT DEVICE IS RECEIVED IN THE FUTURE, THE INVESTIGATION WILL BE REOPENED AND CONDUCTED AS APPROPRIATE. REASON FOR DEVICE EXPLANT AND/OR REOPERATION: DEFLATION. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).

Additional Manufacturer Narrative · 1

AT THE TIME OF THIS REPORT, MENTOR HAS RECEIVED NO INFORMATION REGARDING EXPLANTATION OR AN EXPECTED EXPLANTATION DATE. IT IS UNKNOWN AT THIS TIME IF THE DEVICE WILL BE MADE AVAILABLE FOR RETURN. AS A RESULT, NO PRODUCT FAILURE ANALYSIS CAN BE CONDUCTED, AND NO DETERMINATION OF POSSIBLE CONTRIBUTING FACTORS CAN BE MADE. AS SUCH, THE INVESTIGATION WILL BE CLOSED. IF THE COMPLAINT DEVICE IS RECEIVED IN THE FUTURE, THE INVESTIGATION WILL BE REOPENED AND CONDUCTED AS APPROPRIATE. A MANUFACTURING RECORD EVALUATION IS IN PROGRESS. ONCE COMPLETED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. REASON FOR DEVICE EXPLANT AND/OR REOPERATION: N/A. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT A (B)(6) YEAR-OLD CAUCASIAN FEMALE WHO UNDERWENT PRIMARY BREAST AUGMENTATION WITH A MENTOR SILTEX ROUND MODERATE PROFILE 425CC SALINE PROSTHESIS EXPERIENCED LEFT SIDED DEFLATION POST PROCEDURE. THE DEFLATION WAS DIAGNOSED IN THE PHYSICIAN¿S OFFICE. AT THE TIME OF THIS REPORT, MENTOR HAS RECEIVED NO INFORMATION REGARDING EXPLANTATION OR AN EXPECTED EXPLANTATION DATE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
458601 MENTOR SILTEX ROUND MODERATE PROFILE PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE FWM MENTOR TEXAS 3542670 220652 00081317001829

Patients

Seq Age Sex Outcome Treatment
1 64 YR Other| R