FDA Adverse Event
Malfunction
Summary report: N
RIGHTSPOT PH ENFIT INDICATOR
MDR report key: 11562469
·
Received March 24, 2021
Report
- Report Number
- 11562469
- Event Type
- Malfunction
- Date Received
- March 24, 2021
- Date of Event
- March 2, 2021
- Report Date
- March 15, 2021
- Manufacturer
- RIGHTBIO METRICS
- Product Code
- JLK
- Product Problem
- Yes
- Report Source
- User Facility report
- Reporter Location
- MN
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Description of Event or Problem · 1
NG TUBE INSERTED TO 19CM. RIGHTSPOT PH ENFIT INDICATOR ATTACHED TO CHECK PLACEMENT. WHEN ASPIRATING TO CHECK PLACEMENT, AIR CONTINUED TO COME BUT NO FLUID. WHEN PH INDICATOR DETACHED FROM THE END OF THE NG TUBE AND FEEDING SYRINGE ATTACHED, <1ML AIR WAS ABLE TO BE ASPIRATED. ANOTHER RIGHTSPOT PH INDICATOR WAS OBTAINED AND USED WITH THE SAME RESULTS. PH INDICATORS CHECKED FOR CRACKS, BUT NO CRACKS WERE OBVIOUSLY VISIBLE. NG PLACEMENT DOUBLE CHECKED WITH ANOTHER RN BY AUSCULTATION AND ASPIRATION. BUBBLES AUSCULTATED OVER THE STOMACH AND SMALL AMOUNTS OF BREASTMILK/FORMULA ABLE TO BE ASPIRATED FROM THE NG WITH A FEEDING SYRINGE. RIGHTSPOT PH INDICATOR & PACKAGING SAVED IN A BAG.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 459632 | RIGHTSPOT PH ENFIT INDICATOR | SODIUM HYDROXIDE AND PHENOL RED (TITRIMETRIC), GASTRIC ACIDITY | JLK | RIGHTBIO METRICS | EN0920201 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |