HIGH FREQUENCY VENTED HUMIDIFICATION CHAMBER
Report
- Report Number
- 9611451-2008-00512
- Event Type
- Malfunction
- Date Received
- September 5, 2008
- Report Date
- August 6, 2008
- Manufacturer
- FISHER & PAYKEL HEALTHCARE, LTD.
- Product Code
- BTT
- PMA / PMN Number
- K934140
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NO
- Reporter Occupation
- NURSE
Narratives
THE LOT NUMBERS ASSOCIATED WITH THIS COMPLAINT ARE 050513, 050322 AND 071105 WHICH CORRESPOND WITH DEVICE MFR DATES OF 05/13/2005, 03/22/2005 AND 11/05/2007 RESPECTIVELY. METHOD: THE MR290HFV CHAMBERS WERE VISUALLY EXAMINED FOR CRACKS. RESULTS: ALL THE DEVICES HAVE A VERTICAL CRACK FROM THE CLEFT OF THE FRONT BAFFLE TO THE BASE OF THE DOME. THE CRACKS TO THE CHAMBERS HAVE OCCURRED WHILE IN USE ON THE SENSORMEDICS 3100B VENTILATOR. A LOT CHECK REVEALED NO OTHER COMPLAINTS OF THIS NATURE FOR THIS LOT NUMBER. CONCLUSION: THE CHAMBER CRACKS HAVE MOST LIKELY RESULTED FROM OSCILLATORY STRESSES INDUCED BY THE HIGH FREQUENCY VENTILATOR. DESIGN IMPROVEMENTS WERE MADE TO THE MR290 CHAMBER WHICH RESULTED IN THE MR290HFV VARIANT RECOMMENDED FOR USE IN HIGH FREQUENCY APPLICATIONS. THIS IMPROVEMENT REDUCED THE INCIDENCE OF CRACKING BUT DID NOT ELIMINATE THE PROBLEM ENTIRELY. A CAPA HAS RECENTLY BEEN INITIATED TO FURTHER INVESTIGATE THE PROBLEM OF CRACKED CHAMBERS WHEN USED IN CONJUNCTION WITH HIGH FREQUENCY OSCILLATORY APPLICATIONS, IN PARTICULAR USE OF THE MR290HFV CHAMBER WITH THE SENSORMEDICS 3100B VENTILATOR. OUR MONITORING AND TRENDING OF INCIDENTS INVOLVING CRACKED MR290HFV CHAMBERS DUE TO USE WITH HIGH FREQNUENCY OSCILLATORS HAS A RATE OF OCCURENCE OF 0.022% WORLDWIDE IN THE LAST YEAR TO JULY 2008.
A HOSPITAL IN A FOREIGN COUNTRY REPORTED TO OUR DISTRIBUTOR THAT 3 UNITS OF THE MR290HFV HIGH FREQUENCY AUTOFEED HUMIDIFICATION CHAMBER CRACKED WHEN USED IN CONJUNCTION WITH THE SENSORMEDICS 31009 VENTILATOR. NO PT CONSEQUENCE WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | HIGH FREQUENCY VENTED HUMIDIFICATION CHAMBER | BTT | FISHER & PAYKEL HEALTHCARE, LTD. | MR290HFV |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | SENSORMEDICS 3100B VENTILATOR |