SOCKET WRENCH FOR VEPTR NUT
Report
- Report Number
- 2939274-2021-01494
- Event Type
- Malfunction
- Date Received
- March 23, 2021
- Report Date
- February 24, 2021
- Manufacturer
- WRIGHTS LANE SYNTHES USA PRODUCTS LLC
- Product Code
- HXC
- UDI-DI
- 10705034720128
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CO, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
REPORTER IS A J&J EMPLOYEE. INVESTIGATION SUMMARY. SERVICE AND REPAIR EVALUATION: DURING TESTING AT SERVICE AND REPAIR IT WAS FOUND OUT THAT THE SOCKET WRENCH HAD FAILED IN CALIBRATION. THE REPAIR TECHNICIAN REPORTED THE DEVICE FAILED TORQUE TESTING. THE CAUSE OF THE ISSUE IS FAILED HIGH. THE ITEM WILL BE REPAIRED AND WILL BE RETURNED TO THE CUSTOMER UPON COMPLETION OF THE SERVICE AND REPAIR PROCESS. ATTACHED SERVICE RECORD ROUTER COMPLETED THROUGH OPERATION 20. FINALIZED SERVICE RECORD WILL BE ARCHIVED IN TUNGSTEN DOCUMENT MANAGEMENT SYSTEM. THE EVALUATION WAS CONFIRMED. THE DEVICE WAS DEEMED SERVICEABLE AND WILL BE RETURNED TO THE CUSTOMER, NO DESIGN OR MANUFACTURING ISSUES WERE IDENTIFIED THEREFORE IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED. DEVICE HISTORY LOT. PART NUMBER:03.641.004. SYNTHES LOT NUMBER: J001347-13. SUPPLIER LOT NUMBER: N/A. RELEASE TO WAREHOUSE DATE: DEC 4, 2020. EXPIRATION DATE: N/A. SUPPLIER: (B)(4). NO NCRS WERE GENERATED DURING PRODUCTION. DEVICE HISTORY REVIEW: REVIEW OF THE DEVICE HISTORY RECORD SHOWED THAT THERE WERE NO ISSUES DURING THE MANUFACTURE OF THIS PRODUCT, AND ANY SUBCOMPONENTS, WHICH WOULD CONTRIBUTE TO THIS COMPLAINT CONDITION. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
IT WAS REPORTED ON (B)(6), 2021, DURING TESTING AT SERVICE AND REPAIR IT WAS FOUND OUT THAT THE SOCKET WRENCH HAD FAILED IN CALIBRATION. THERE WAS NO PATIENT INVOLVEMENT. THIS REPORT IS FOR ONE (1) SOCKET WRENCH FOR VEPTR NUT. THIS IS REPORT 1 OF 1 FOR (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 445865 | SOCKET WRENCH FOR VEPTR NUT | WRENCH | HXC | WRIGHTS LANE SYNTHES USA PRODUCTS LLC | 03.641.004 | J001347-13 | 10705034720128 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |