IRRIGATING DISPOSABLE BIPOLAR FORCEPS 8 X 0.5MM
Report
- Report Number
- 11553289
- Event Type
- Malfunction
- Date Received
- March 23, 2021
- Date of Event
- March 3, 2021
- Report Date
- March 8, 2021
- Manufacturer
- KATALYST SURGICAL LLC
- Product Code
- GEI
- UDI-DI
- 00817489027202
- Product Problem
- Yes
- Report Source
- User Facility report
- Reporter Location
- VA
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
CODMAN IRRIGATING DISPOSABLE BIPOLAR FORCEPS WERE FOUND TO HAVE A 3M INDICATOR/ INSTRUMENT PROTECTOR IN THEM WHERE THE ETHYLENE OXIDE INDICATOR DID NOT CHANGE ON SOME OF THE FORCEPS PACKAGING. CONTACTED CODMAN REP ABOUT THIS ISSUE. THEIR SUPERIOR AND QUALITY DEPARTMENT STATED THE 3M INDICATOR/INSTRUMENT PROTECTOR WERE ONLY PLACE HOLDERS. THE INDICATOR THAT IS USED TO DETERMINE STERILITY IS PLACED ON THE EXTERIOR OF THE CORRUGATED SHIPPING BOX. THE STICKER ON THE OUTSIDE IS PURPLE AND WILL TURN GREEN WHEN IN CONTACT WITH THE ETHYLENE OXIDE (EO). THIS DOT INDICATOR IS USED TO INDICATE STERILIZATION BY EO FOR THESE DISPOSABLES. THERE ARE 5 BIPOLAR FORCEPS (DOUBLE PEEL PACKAGES FOR EACH DISPOSABLE FORCEP) IN A WHITE CARDBOARD BOX AND SEVERAL WHITE CARDBOARD BOXES ARE PACKAGED WITHIN THE CORRUGATED SHIPPING BOX. MULTIPLE REF#S AND LOT#S INVOLVED - 3 EXAMPLES ADDED TO THIS EVENT. THE 3M INDICATOR/INSTRUMENT PROTECTOR HAS 2 YELLOW DIAMONDS FOR EO THAT WILL CHANGE TO RUST/RED IF EXPOSED TO EO APPROPRIATELY. SOME OF THE 3M INDICATORS CHANGED APPROPRIATELY AND SOME REMAINED YELLOW. WE QUESTION IF THE EO GAS EFFECTIVELY PENETRATED THROUGH ALL PACKAGING TO 3M INDICATOR AND IF CONTENTS ARE STERILE. MULTIPLE BOXES OF FORCEPS WERE RETURNED TO CODMAN FOR ANALYSIS. FEEDBACK FROM (B)(4) QUALITY CONTROL STATED THE 3M INDICATOR/ INSTRUMENT PROTECTOR IS ONLY USED AS AN INSTRUMENT PROTECTOR/HOLDER AND THEY NOTED SOME OF THE YELLOW DIAMOND INDICATORS THAT HAD TURNED SLIGHTLY GOLD WERE DETERMINED TO BE RUST IN COLOR AND THEREFORE, STERILE. PICTURES X 4 ATTACHED: #1 PICTURE OF UNCHANGED INDICATOR WITHIN DOUBLE PEEL PACKAGING NEXT TO INDICATOR THAT TURNED RUST/RED AS APPROPRIATE FOR EO STERILIZATION. #2 PICTURE OF CLOSE UP OF UNCHANGED INDICATOR WITHIN PACKAGING. #3 PICTURE OF NEW UNUSED 3M INDICATOR - NOTE YELLOW COLORED DIAMONDS. #4 PICTURE OF USED/CONVERTED 3M INDICATOR - NOTE RUST/RED COLORED DIAMONDS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 451227 | IRRIGATING DISPOSABLE BIPOLAR FORCEPS 8 X 0.5MM | ELECTROSURGICAL, CUTTING COAGULATION ACCESSORIES | GEI | KATALYST SURGICAL LLC | 9008050IR | M16373 | 00817489027202 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |