FDA Adverse Event Malfunction Summary report: N

PROPEL

MDR report key: 11553059 · Received March 23, 2021

Report

Report Number
11553059
Event Type
Malfunction
Date Received
March 23, 2021
Date of Event
February 21, 2021
Report Date
March 16, 2021
Manufacturer
INTERSECT ENT, INC.
Product Code
OWO
Product Problem
Yes
Report Source
User Facility report
Reporter Location
IL
Reporter Occupation
NURSE

Narratives

Description of Event or Problem · 1

YOUNG ADULT FEMALE WAS HAVING AN OUTPATIENT PROCEDURE TO TREAT CHRONIC SINUSITIS. WHILE USING THE PROPEL SINUS IMPLANT 23MM DURING THE CASE, DURING THE DEPLOYMENT PROCESS, THE SPINDLES ON THE IMPLANT WERE NOT RELEASING FROM THE DEPLOYMENT DEVICE. THE PUSHER INSIDE THE DEPLOYMENT PIECE APPEARED TO BE PINCHING THE IMPLANT AND NOT ALLOWING IT TO RELEASE. THE SURGEON WAS ABLE TO GET THE IMPLANT OUT OF THE DELIVERY SYSTEM, BUT THAT PROCESS AFFECTED THE IMPLANT ITSELF. THIS ERROR OCCURRED WITH TWO DIFFERENT DEVICES. THIS IS A SINGLE-USE, DISPOSABLE ITEM. THERE WAS NO HARM TO THE PATIENT. THE SURGEON FELT THERE WAS AN ISSUE WITH DEPLOYMENT DUE TO MANUFACTURER ERROR WITH THE DEVICE, BUT THE PRODUCT REPRESENTATIVE FELT IT WAS POTENTIALLY OPERATOR ERROR WHEN LOADING THE PROPEL STENT IN THE DEVICE. PLAN IS TO RE-TRAIN STAFF.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
449931 PROPEL DRUG-ELUTING SINUS STENT OWO INTERSECT ENT, INC. 70011 90920001 AND 91106003

Patients

Seq Age Sex Outcome Treatment
1 13140 DA