FDA Adverse Event Malfunction Summary report: N

BD VACUTAINER® SERUM BLOOD COLLECTION TUBES

MDR report key: 11543527 · Received March 22, 2021

Report

Report Number
1024879-2021-00199
Event Type
Malfunction
Date Received
March 22, 2021
Date of Event
March 1, 2021
Report Date
August 10, 2022
Manufacturer
BECTON, DICKINSON & CO., (BD)
Product Code
JKA
UDI-DI
50382903678124
PMA / PMN Number
BK050036
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

H6: INVESTIGATION SUMMARY BD RECEIVED ONE (1) SAMPLE AND ONE (1) PHOTO FROM MATERIAL # 367812 TUBE, BATCH # 0100893 FOR INVESTIGATION. THE PHOTO WAS REVIEWED AND THE CUSTOMER¿S INDICATED FAILURE MODE FOR FOREIGN MATTER WITH THE INCIDENT LOT WAS OBSERVED. ADDITIONALLY, THE CUSTOMER SAMPLE WAS EVALUATED BY VISUAL EXAMINATION AND THE INDICATED FAILURE MODE FOR FOREIGN MATTER WITH THE INCIDENT LOT WAS OBSERVED. BD DETERMINED THAT THE ROOT CAUSE OF THE FOREIGN MATTER EMBEDDED IN THE TUBE (SPOTS IN TUBE) WAS ATTRIBUTED TO THE INJECTION MOLDING START-UP PROCESS, WITH INADEQUATE PURGING OF THE LINE OCCURRING. BASED ON A REVIEW OF THE DEVICE HISTORY RECORD FOR THE INCIDENT LOT, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET. THERE WERE NO RELATED QUALITY ISSUES DURING MANUFACTURING OF THE PRODUCT.

Description of Event or Problem · 0

IT WAS REPORTED WHEN USING THE BD VACUTAINER® SERUM BLOOD COLLECTION TUBES THERE WAS FOREIGN MATTER IN THE TUBE(S) BIOLOGICAL AND NON-BIOLOGICAL. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: TRANSLATED TO ENGLISH. THE CUSTOMER STATED: BEFORE USE, THERE WERE MANY BLACK SPOTS FOUND IN THE TUBE."

Additional Manufacturer Narrative · 1

INVESTIGATION SUMMARY: BD HAD NOT RECEIVED SAMPLES, BUT (1) ONE PHOTO WAS PROVIDED FOR INVESTIGATION. THE PHOTO WAS REVIEWED AND THE INDICATED FAILURE MODE FOR FOREIGN MATTER WITH THE INCIDENT LOT WAS OBSERVED. THE EXACT IDENTITY OF THE FOREIGN MATTER COULD NOT BE DETERMINED FROM THE PHOTO ALONE. THE PHYSICAL SAMPLE WOULD BE NEEDED FOR FURTHER IDENTIFICATION. BASED ON A REVIEW OF THE DEVICE HISTORY RECORD FOR THE INCIDENT LOT, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET. THERE WERE NO RELATED QUALITY ISSUES DURING MANUFACTURING OF THE PRODUCT.

Description of Event or Problem · 1

IT WAS REPORTED WHEN USING THE BD VACUTAINER® SERUM BLOOD COLLECTION TUBES THERE WAS FOREIGN MATTER IN THE TUBE(S) BIOLOGICAL AND NON-BIOLOGICAL. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: TRANSLATED TO ENGLISH. THE CUSTOMER STATED: BEFORE USE, THERE WERE MANY BLACK SPOTS FOUND IN THE TUBE."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
436309 BD VACUTAINER® SERUM BLOOD COLLECTION TUBES BLOOD SPECIMEN COLLECTION DEVICE JKA BECTON, DICKINSON & CO., (BD) 367812 0100893 50382903678124

Patients

Seq Age Sex Outcome Treatment
1 Unknown