FDA Adverse Event
Malfunction
Summary report: N
PICO70
MDR report key: 11542063
·
Received March 22, 2021
Report
- Report Number
- 3002807968-2021-00008
- Event Type
- Malfunction
- Date Received
- March 22, 2021
- Date of Event
- March 3, 2021
- Report Date
- March 22, 2021
- Manufacturer
- RADIOMETER MEDICAL APS
- Product Code
- JKA
- PMA / PMN Number
- K962158
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- DA
- Reporter Occupation
- 003
Narratives
Description of Event or Problem · 1
RADIOMETER HAS BECOME AWARE OF AN ISSUE WITH PICO70 ARTERIAL BLOOD SAMPLER. THE ISSUE RELATES TO THE STERILE BARRIER SYSTEM FOR THE PRODUCT. DURING INTERNAL STRESS TESTING ON PICO70 ARTERIAL BLOOD SAMPLERS A BREACH IN THE STERILE BARRIER SYSTEM WAS OBSERVED IN ONE SAMPLE .THE PROBLEM HAS ONLY BEEN OBSERVED INTERNALLY DURING TESTING. RADIOMETER HAS ON 2021-03-18 INITIATED A RECALL WITH RADIOMETER REF. FAN 915-415 OF ALL PICO70 BLOOD SAMPLERS. A CORRECTION AND REMOVAL REPORT WILL BE SUBMITTED TO THE FDA.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 435532 | PICO70 | PICO70 | JKA | RADIOMETER MEDICAL APS | 956-519, 956-522 & 956-533 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |