FDA Adverse Event Injury Summary report: N

CONTOUR NEXT

MDR report key: 11541946 · Received March 22, 2021

Report

Report Number
1810909-2021-00142
Event Type
Injury
Date Received
March 22, 2021
Date of Event
February 25, 2021
Report Date
February 25, 2021
Manufacturer
ASCENSIA DIABETES CARE US INC.
Product Code
NBW
PMA / PMN Number
K160682
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

THE CUSTOMER RETURNED THE SUSPECTED CONTOUR NEXT (CONNECTED) METER FOR EVALUATION. NO TEST STRIPS WERE RETURNED. THEREFORE, THE RETURNED METER WAS TESTED WITH THE IN-HOUSE CONTOUR NEXT TEST STRIPS FROM LOT # 9CPEG18B USING A BLOOD SAMPLE, WHICH GAVE A SATISFACTORY PERFORMANCE. SECTION D4 HAS BEEN UPDATED WITH THE CORRECT LOT #.

Additional Manufacturer Narrative · 1

THE PATIENT/FAMILY WAS THE INITIAL REPORTER, SO PERSONAL INFORMATION WAS NOT ENTERED. NO INFORMATION WAS CAPTURED AS THE CUSTOMER'S AGE AND WEIGHT WERE NOT PROVIDED. THE MODEL # WAS NOT PROVIDED. THE CONTOUR NEXT (CONNECTED) METER IS NOT MARKETED IN THE UNITED STATES. HOWEVER, THE DEVICE IS SIMILAR TO THE CONTOUR NEXT ONE METER WITH THE PRODUCT CODE NBW AND 510 (K) # K160682, WHICH IS MARKETED IN THE UNITED STATES.

Description of Event or Problem · 1

AN ADVOCATE FROM (B)(6) CALLED ON BEHALF OF HIS FATHER TO REPORT THAT WITHIN 5 MINUTES, THEY OBTAINED BLOOD GLUCOSE READINGS OF 250 MG/DL AND 150 MG/DL WITH THEIR CONTOUR NEXT (CONNECTED) METER. THE CUSTOMER WAS EXPERIENCING SYMPTOMS OF HYPOGLYCEMIA SUCH AS TREMBLING, DIZZINESS AND FOAMING AT THE MOUTH. THE CUSTOMER BECAME UNRESPONSIVE. AN AMBULANCE WAS CALLED AND 15 MINUTES LATER, A READING OF 19 MG/DL WAS RECEIVED WITH AN ALTERNATE METER. THE CUSTOMER WAS GIVEN AN UNKNOWN INFUSION AFTER WHICH HE RECOVERED WELL. THE ADVOCATE WAS ADVISED TO RETURN THE DEVICE FOR EVALUATION. REPLACEMENT METER KIT AND TEST STRIPS WERE SENT TO THE CUSTOMER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
438749 CONTOUR NEXT BLOOD GLUCOSE TEST STRIPS NBW ASCENSIA DIABETES CARE US INC. 9CPEG18B

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention