FDA Adverse Event Injury Summary report: N

STERICHEK TOTAL CHLORAMINE AND RESIDUAL CHLORINE

MDR report key: 1153743 · Received September 3, 2008

Report

Report Number
1153743
Event Type
Injury
Date Received
September 3, 2008
Date of Event
July 14, 2008
Report Date
September 2, 2008
Manufacturer
HACH COMPANY
Product Code
MSY
Adverse Event
Yes
Product Problem
Yes
Report Source
User Facility report
Reporter Location
CA, US
Reporter Occupation
RISK MANAGER

Narratives

Description of Event or Problem · 1

MACHINE IN USE SEVERAL TIMES. WAS CLEANED/REVISED AND TEST FOR BLEACH SHOWED. ON 7/14 PATIENT CONDITION WORSENED WITHIN 10 MINUTES OF START OF DIALYSIS. AFTER RINSEBACK BLOOD LINE TURNED LIGHT BROWN. DRAIN LINE NEGATIVE FOR BLEACH; BLUE HANSON CONNECTOR AND I PRESSURE BLEACH. PATIENT TO ED THEN TCN FOR DIALYSIS, AND OVERNIGHT STAY. OUR FACILITY NOTIFIED OF HACH REAGENT STRIP RECALL ON 7/24 VIA A RENAL NETWORK. FALSE NEGATIVES AND POSITIVES BEING FOUND, THE REPORT SHOWS. NOT CLEAR WHAT CAUSE OF PATIENT'S CHANGE IN CONDITION. REAGENT STRIPS FAULTY. HOWEVER BLEACH EXPOSURE IS AMONG THE POSSIBILITIES. INCREASED LENGTH OF TUBING TO RINSE ALSO. DIALYSIS EQUIPMENT MANUFACTURERS NOTIFIED OF INCIDENT AND RESPONDED (FRESENIUS) CONNECTOR IS TWO STEP TO CLICK/SNAP. ALSO, POTENTIAL FOR BLEACH TO ACCUMULATE DURING "LAST PEDOWN".

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 STERICHEK TOTAL CHLORAMINE AND RESIDUAL CHLORINE REAGENT STRIPS MSY HACH COMPANY 811909
2 ASSOCIATED EQUIPMENT FRESENIUS DIALYSIS EQUIPMENT FII FRESENIUS MEDICAL CARE

Patients

Seq Age Sex Outcome Treatment
1 64 YR Other