FDA Adverse Event Malfunction Summary report: N

9900

MDR report key: 1153645 · Received September 11, 2008

Report

Report Number
1720753-2008-22278
Event Type
Malfunction
Date Received
September 11, 2008
Date of Event
May 28, 2008
Report Date
June 6, 2008
Manufacturer
GE OEC MEDICAL SYSTEMS INC.
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE GE SERVICE REP RECALIBRATED THE SYSTEM MAG MODE TO BE WITHIN SPEC. SYSTEM FUNCTIONED AS INTENDED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THE BEAM WAS NOT RESTRICTED TO INPUT PHOPHOR + OR - 3% SID SINGLE DIRECTION AND + OR - 4% SID COMBINED DIRECTION IN MAG 2" MODE. NO PT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 9900 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS INC. NA NA

Patients

Seq Age Sex Outcome Treatment
1