PUMP IN STYLE RETAIL EN FR ES (YELLOWFIN)
Report
- Report Number
- 1419937-2021-00022
- Event Type
- Injury
- Date Received
- March 19, 2021
- Date of Event
- February 24, 2021
- Manufacturer
- MEDELA LLC
- Product Code
- HGX
- UDI-DI
- 00020451401454
- PMA / PMN Number
- K200508
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- OTHER
Narratives
CUSTOMER SERVICE CONDUCTED TROUBLESHOOTING WITH THE CUSTOMER, INCLUDING TRYING TO POWER THE PUMP ON AND TRYING MULTIPLE OUTLETS. THE CUSTOMER WAS SENT A REPLACEMENT DEVICE AND RETURN OF HER ORIGINAL DEVICE WAS REQUESTED FOR TESTING/EVALUATION. THE CUSTOMER WAS CONTACTED BY A COMPLAINT HANDLER ON MULTIPLE OCCASIONS, INCLUDING IN WRITING, TO GET ADDITIONAL INFORMATION, WITH NO RESPONSE AS OF THE DATE OF THIS REPORT. BASED ON THE RESULTS OF CA11-001, IT CANNOT BE DEFINITIVELY CONCLUDED THAT THE PUMP CAUSED OR CONTRIBUTED TO THE CUSTOMER'S MASTITIS. THE ESTIMATED INCIDENCE OF MASTITIS IN LACTATING WOMEN, WHETHER USING A BREAST PUMP OR NOT, ACCORDING TO PUBLISHED CLINICAL LITERATURE CAN BE AS HIGH AS 33%. IN FACT, CLINICAL GUIDELINES SUGGEST THE USE OF A BREAST PUMP TO FACILITATE WITHDRAWAL OF BREAST MILK DURING BOUTS OF MASTITIS. THE COMPLAINT RATE OF MASTITIS ACROSS ALL REPORTED FAILURES, ACROSS ALL MEDELA BREAST PUMPS, IS 0.008% FOR THE PERIOD OF JANUARY 2013 TO AUGUST 2017. MASTITIS IS USUALLY A BENIGN, SELF-LIMITING INFECTION WITH FEW CONSEQUENCES FOR THE SUCKLING INFANT. THE RISK OF MASTITIS IS HIGHER AMONG WOMEN WHO HAVE BREASTFED PREVIOUSLY, ESPECIALLY THOSE WITH A HISTORY OF MASTITIS." RIORDAN & WAMBACH, 4TH ED. P. 294: BREASTFEEDING AND HUMAN LACTATION. MASTITIS REQUIRES PROMPT MEDICAL ATTENTION FOR THE MOTHER FOR PAIN RELIEF AND PRESCRIPTION ANTIBIOTICS TO AVOID PROGRESSION TO OVERWHELMING SEPSIS.
ON (B)(6) 2021, THE CUSTOMER ALLEGED TO MEDELA LLC THAT HER PUMP IN STYLE MAXFLOW BREAST PUMP WAS POWERING OFF ON ITS OWN AFTER TWO MINUTES. SHE ADDITIONALLY ALLEGED THAT SHE HAD TO GO TO THE HOSPITAL BECAUSE SHE HAD MASTITIS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 419927 | PUMP IN STYLE RETAIL EN FR ES (YELLOWFIN) | PUMP, BREAST, POWERED | HGX | MEDELA LLC | 101041361 | 0000017580 | 00020451401454 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |