FDA Adverse Event Malfunction Summary report: N

GE OEC 9800

MDR report key: 1152945 · Received September 10, 2008

Report

Report Number
1720753-2008-22259
Event Type
Malfunction
Date Received
September 10, 2008
Date of Event
May 27, 2008
Report Date
June 6, 2008
Manufacturer
GE OEC MEDICAL SYSTEMS INC.
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE REPRESENTATIVE EVALUATED THE SYSTEM AND REPLACED THE SNUBBER BOARD AND RE-GREASED THE CANDLESTICK CONNECTORS. SYSTEM OPERATES AS INTENDED.

Description of Event or Problem · 1

CUSTOMER REPORTED NO IMAGE ON MONITOR DURING FLUORO. NO PATIENT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 GE OEC 9800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS INC. 9800 NA

Patients

Seq Age Sex Outcome Treatment
1