SYRINGE 10ML LL BNS
Report
- Report Number
- 1213809-2021-00148
- Event Type
- Malfunction
- Date Received
- March 19, 2021
- Date of Event
- February 22, 2021
- Report Date
- March 29, 2021
- Manufacturer
- BECTON DICKINSON MEDICAL SYSTEMS
- Product Code
- FMF
- PMA / PMN Number
- K980987
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
H.6. INVESTIGATION: ONE PHOTO OF A LOOSE 10ML SYRINGE WAS RECEIVED AND EVALUATED. IT WAS OBSERVED THERE WAS SMALL AREAS OF MISSING PRINT NEAR THE 9ML MARKING. THE MISSING PRINT HAD A DRAGGED OR SCRAPED APPEARANCE. NO SINGLE ITEM WAS MISSING MORE THAN 50% WHICH WAS ACCEPTABLE PER PRODUCT SPECIFICATION. POTENTIAL ROOT CAUSE FOR THE MISSING PRINT DEFECT MAY BE ASSOCIATED WITH THE ASSEMBLY PROCESS. IT IS POSSIBLE THERE WAS A SLIGHT MISFEED CAUSING THE PRINT TO BE RUBBED OFF. A DEVICE HISTORY RECORD REVIEW SHOWED NO REJECTED INSPECTIONS OR QUALITY ISSUES DURING THE PRODUCTION OF THE PROVIDED LOT NUMBER THAT COULD HAVE CONTRIBUTED TO THE REPORTED DEFECT. H3 OTHER TEXT : SEE H.10.
IT WAS REPORTED THAT SYRINGE 10ML LL BNS HAD SCALE MARKING ISSUES. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: MATERIAL NO: 30102, BATCH NO: 9190569. IT WAS REPORTED SYRINGE WITH BLURRED LETTERS.
DATE OF EVENT: UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED, BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.
IT WAS REPORTED THAT SYRINGE 10ML LL BNS HAD SCALE MARKING ISSUES. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: MATERIAL NO: 301029, BATCH NO: 9190569. IT WAS REPORTED SYRINGE WITH BLURRED LETTERS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 428882 | SYRINGE 10ML LL BNS | PISTON SYRINGE | FMF | BECTON DICKINSON MEDICAL SYSTEMS | 9190569 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |