MINIMED 770G US SYSTEM BLE CONNECT 3.0 MG/DL
Report
- Report Number
- 2032227-2021-126320
- Event Type
- Death
- Date Received
- March 19, 2021
- Date of Event
- February 21, 2021
- Report Date
- February 7, 2022
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OZP
- UDI-DI
- 000000763000414344
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
- Health Professional
- N
Narratives
ON (B)(6) 2021 THE INSULIN PUMP WAS RETURNED DUE TO CUSTOMER PASSING AWAY. CUSTOMER'S HUSBAND ALLEGES PUMP BEING DEFECTIVE. THE DEVICE DID NOT HAVE A BATTERY INSTALLED WHEN RECEIVED. DEVICE PASSED THE FUNCTIONAL TEST, INCLUDING THE SELF TEST, SLEEP CURRENT MEASUREMENT, ACTIVE CURRENT MEASUREMENT, REWIND TEST, PRIME/SEATING TEST, BASIC OCCLUSION TEST, OCCLUSION TEST, FORCE SENSOR TEST, DISPLACEMENT TEST AND THE DELIVERY ACCURACY TEST AT 0.0873 INCHES. NO OVER DELIVERY ANOMALY OR UNDER DELIVERY ANOMALY NOTED. SUCCESSFULLY DOWNLOADED HISTORY FILES AND TRACES USING THUMP. SUCCESSFULLY UPLOADED TO CARELINK AND GENERATED REPORTS. THE TEST P-CAP AND RESERVOIR DOES LOCK IN PLACE IN THE RESERVOIR COMPARTMENT. DEVICE HAD SCRATCHED CASE AND A PILLOWING KEYPAD OVERLAY. DATA ANALYSIS: (B)(6) 2021 DAILYTOTALOFALLINSULINDELIVERED: 279500 (27.95 U) (B)(6) 2021 DAILYTOTALOFALLINSULINDELIVERED: 324250 (32.425 U) (B)(6) 2021 DAILYTOTALOFALLINSULINDELIVERED: 283500 (28.35 U) 02/18/2021 DAILYTOTALOFALLINSULINDELIVERED: 300500 (30.05 U) 02/19/2021 DAILYTOTALOFALLINSULINDELIVERED: 356750 (35.675 U) (B)(6) 2021 DAILYTOTALOFALLINSULINDELIVERED: 287750 (28.775 U) (B)(6) 2021 DAILYTOTALOFALLINSULINDELIVERED: 200750 (20.075 U). ALLEGATION FOR INSULIN PUMP BEING DEFECTIVE NOT CONFIRMED DURING FULL FUNCTIONAL TESTING. THIS MDR RELATED TO THE PUERTO RICO MANUFACTURING SITE HAS BEEN ASSIGNED A MEDWATCH NUMBER FROM THE MEDTRONIC MINIMED NORTHRIDGE SITE, PER VARIANCE 5. MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS" OR HAS "MALFUNCTIONED". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT.
DEVICE DID NOT HAVE A BATTERY INSTALLED WHEN RECEIVED. DEVICE PASSED THE FUNCTIONAL TEST, INCLUDING THE SELF TEST, SLEEP CURRENT MEASUREMENT, ACTIVE CURRENT MEASUREMENT, REWIND TEST, PRIME/SEATING TEST, BASIC OCCLUSION TEST, OCCLUSION TEST, FORCE SENSOR TEST, DISPLACEMENT TEST AND THE DELIVERY ACCURACY TEST AT 0.0873 INCHES. NO OVER DELIVERY ANOMALY OR UNDER DELIVERY ANOMALY NOTED. DEVICE UPLOADED PROPERLY USING CARE LINK. DEVICE HAD SCRATCHED CASE AND A PILLOWING KEYPAD OVERLAY. THE TEST P-CAP AND RESERVOIR DOES LOCK IN PLACE IN THE RESERVOIR COMPARTMENT. (B)(4). MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS" OR HAS "MALFUNCTIONED". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT.
(B)(4). CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. WE THEREFORE CONSIDER THIS REPORT COMPLETE TO THE BEST OF OUR KNOWLEDGE.
IT WAS REPORTED VIA PHONE CALL THAT THE CUSTOMER PASSED AWAY AT HOME. THE CAUSE OF DEATH WAS DIABETIC COMPLICATIONS. THE CALLER STATED THAT THE CUSTOMER HAD NO OTHER ILLNESSES THAT MAY HAVE LED TO THE CUSTOMER'S PASSING. THE CUSTOMERS BLOOD GLUCOSE WAS UNKNOWN AT THE TIME OF DEATH. THE CUSTOMER WAS WEARING THE INSULIN PUMP AT THE TIME OF DEATH. IT IS UNKNOWN IF THE AUTO MODE ON THE INSULIN PUMP WAS ACTIVE AND AUTOMATICALLY ADJUSTING INSULIN DELIVERY DURING THE EVENT. THE CUSTOMER WAS USING SENSORS AT THE TIME OF THE INCIDENT. THE CALLER STATED THAT AT THE ONSET OF THE ILLNESS, THE CUSTOMER WAS VOMITING REALLY BADLY AND WAS SLUGGISH THE NEXT DAY, SO THEY MADE SOME ADJUSTMENTS ON THEIR INSULIN PUMP. THE CALLER DECLINED TO RETURN THE INSULIN PUMP FOR ANALYSIS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 425117 | MINIMED 770G US SYSTEM BLE CONNECT 3.0 MG/DL | ARTIFICIAL PANCREAS DEVICE SYSTEM, SINGLE HORMONAL CONTROL | OZP | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-1880 | HG4XBJ9ZZ | 000000763000414344 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 56 YR | Female | Death | FRN-MMT-332-RSVR,UNOMED SET| FRN-MMT-332-RSVR,UNOMED SET |