FDA Adverse Event Malfunction Summary report: N

VDW SILVER RECIPROCATING MOTOR

MDR report key: 11522067 · Received March 18, 2021

Report

Report Number
9611053-2021-00079
Event Type
Malfunction
Date Received
March 18, 2021
Report Date
March 18, 2021
Manufacturer
DENTSPLY VDW GMBH
Product Code
EBW
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
DENTIST

Narratives

Additional Manufacturer Narrative · 1

WHILE NO SERIOUS INJURY RESULTED IN THIS EVENT, CALIBRATION RELATED ISSUES IN ENDO MOTORS HAVE LED TO FILE SEPARATION IN THE PAST. THEREFORE, IT MUST BE PRESUMED THAT RECURRENCE OF THIS MALFUNCTION COULD POSSIBLY CAUSE OR CONTRIBUTE TO A SERIOUS INJURY OR REQUIRE MEDICAL OR SURGICAL INTERVENTION TO PRECLUDE SUCH. AS SUCH, THIS EVENT IS REPORTABLE PER 21CFR PART 803. CONTRA ANGLE 141669 (2019) (NECK DRIVE LOOSE) WAS ATTACHED. MICROMOTOR SMR2081405 AND FOOT CONTROL 186193 TESTED, WITHOUT ERRORS.

Description of Event or Problem · 1

IN THIS EVENT IT WAS REPORTED THAT A VDW.SILVER RECIPROC CAUSES ERROR CODE 2 AT CALIBRATION. NO INJURY OCCURRED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
411297 VDW SILVER RECIPROCATING MOTOR CONTROLLER, FOOT, HANDPIECE AND CORD EBW DENTSPLY VDW GMBH NA NA

Patients

Seq Age Sex Outcome Treatment
1