FDA Adverse Event
Malfunction
Summary report: N
VDW SILVER RECIPROCATING MOTOR
MDR report key: 11522067
·
Received March 18, 2021
Report
- Report Number
- 9611053-2021-00079
- Event Type
- Malfunction
- Date Received
- March 18, 2021
- Report Date
- March 18, 2021
- Manufacturer
- DENTSPLY VDW GMBH
- Product Code
- EBW
- PMA / PMN Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- DENTIST
Narratives
Additional Manufacturer Narrative · 1
WHILE NO SERIOUS INJURY RESULTED IN THIS EVENT, CALIBRATION RELATED ISSUES IN ENDO MOTORS HAVE LED TO FILE SEPARATION IN THE PAST. THEREFORE, IT MUST BE PRESUMED THAT RECURRENCE OF THIS MALFUNCTION COULD POSSIBLY CAUSE OR CONTRIBUTE TO A SERIOUS INJURY OR REQUIRE MEDICAL OR SURGICAL INTERVENTION TO PRECLUDE SUCH. AS SUCH, THIS EVENT IS REPORTABLE PER 21CFR PART 803. CONTRA ANGLE 141669 (2019) (NECK DRIVE LOOSE) WAS ATTACHED. MICROMOTOR SMR2081405 AND FOOT CONTROL 186193 TESTED, WITHOUT ERRORS.
Description of Event or Problem · 1
IN THIS EVENT IT WAS REPORTED THAT A VDW.SILVER RECIPROC CAUSES ERROR CODE 2 AT CALIBRATION. NO INJURY OCCURRED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 411297 | VDW SILVER RECIPROCATING MOTOR | CONTROLLER, FOOT, HANDPIECE AND CORD | EBW | DENTSPLY VDW GMBH | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |